5
6
Flicker
IEC 6100-3-3
Complies
RF Emissions
CISPR 11
Class B
The BT-2.4-WS is suitable for use in all establishments,
including domestic, and those directly connected to the public
low-voltage power supply network that supplies buildings
used for domestic purposes.
RF Emissions
CISPR 11
Group 1
The BT-2.4-WS uses RF energy only for its inernal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
The BT-2.4-WS (SenseWear
®
biotransceiver) is intended for use in the electromagnetic environment specifi ed below.
The customer or user of the BT-2.4-WS should ensure that it is used in such an environment.
Guidance and Manufacturer’s Declaration - Emissions
Emissions Test
Compliance
Electromagnetic Environment - Guidance
Important Information About the SenseWear
®
biotransceiver
Computer and Wireless Communicator
1.5m
Diagram not to scale.
1.5m
1.5m
2.5m
Patient Environment
Harmonics
IEC 6100-3-2
Class A
3A/m
Power frequency magnetic fi elds should be that
of a typical commercial or hospital environment.
Power
Frequency 50/60Hz
Magnetic Field
IEC 61000-4-8
3A/m
N/A
N/A
Voltage Dips/
Dropout
IEC 61000-4-11
>95% Dip for
0.5 Cycles
60% Dip for 5 Cycles
30% Dip for 25 Cycles
>95% Dip for 5
Seconds
N/A
Surge
IEC 61000-4-5
±1kV Differential
±2kV Common
N/A
N/A
N/A
EFT
IEC 61000-4-4
±2kV Mains
±1kV I/Os
±6kV Contact
±8kV Air
Floors should be wood, concrete, or ceramic
tile. If fl oors are synthetic, the r/h should be at
least 30%.
ESD
IEC 61000-4-2
±6kV Contact
±8kV Air
The BT-2.4-WS (SenseWear
®
biotransceiver) is intended for use in the electromagnetic environment specifi ed below.
The customer or user of the BT-2.4-WS should ensure that it is used in such an environment.
Guidance and Manufacturer’s Declaration - Immunity
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment - Guidance
Important Information About the SenseWear
®
biotransceiver