Flicker
IEC 6100-3-3

Complies

RF Emissions
CISPR 11

Class B

The BT-2.4-WS is suitable for use in all establishments, 
including domestic, and those directly connected to the public 
low-voltage power supply network that supplies buildings 
used for domestic purposes.

RF Emissions 
CISPR 11 

Group 1

The BT-2.4-WS uses RF energy only for its inernal function. 
Therefore, its RF emissions are very low and are not likely to 
cause any interference in nearby electronic equipment.

The BT-2.4-WS (SenseWear

®

 biotransceiver) is intended for use in the electromagnetic environment specifi ed below. 

The customer or user of the BT-2.4-WS should ensure that it is used in such an environment. 

Guidance and Manufacturer’s Declaration - Emissions

Emissions Test

Compliance

Electromagnetic Environment - Guidance

Important Information About the SenseWear

®

 biotransceiver

Computer and Wireless Communicator

1.5m

Diagram not to scale.

1.5m

1.5m

2.5m

Patient Environment 

Harmonics 
IEC 6100-3-2

Class A

3A/m

Power frequency magnetic fi elds should be that 
of a typical commercial or hospital environment.

Power 
Frequency 50/60Hz

Magnetic Field
IEC 61000-4-8

3A/m

N/A

N/A

Voltage Dips/
Dropout
IEC 61000-4-11

>95% Dip for
0.5 Cycles

60% Dip for 5 Cycles

30% Dip for 25 Cycles

>95% Dip for 5 
Seconds

N/A

Surge
IEC 61000-4-5

±1kV Differential
±2kV Common

N/A

N/A

N/A

EFT
IEC 61000-4-4

±2kV Mains
±1kV I/Os

±6kV Contact
±8kV Air

Floors should be wood, concrete, or ceramic 
tile. If fl oors are synthetic, the r/h should be at 
least 30%.

ESD
IEC 61000-4-2

±6kV Contact
±8kV Air

The BT-2.4-WS (SenseWear

®

 biotransceiver) is intended for use in the electromagnetic environment specifi ed below.  

The customer or user of the BT-2.4-WS should ensure that it is used in such an environment. 

Guidance and Manufacturer’s Declaration - Immunity

Immunity Test

IEC 60601

Test Level

Compliance

Level

Electromagnetic

Environment - Guidance

Important Information About the SenseWear

®

 biotransceiver

Содержание SenseWear

Страница 1: ...11th Floor Pittsburgh PA 15222 USA P 412 288 9901 F 412 288 9902 www bodymedia com BodyMedia International Srl Via G Balzaretti 15 20133 Milan Italy P 39 02 2774121 F 39 02 27741250 www bodymedia eu b...

Страница 2: ...ts for a safe medical device under applicable directives However this product alone is not meant to substitute for proper medical diagnosis care or treatment Any decisions based on the data from this...

Страница 3: ...teries Batteries may explode or leak and can cause burn injury if recharged disposed of in fire or disassembled Dispose of properly For further information about the disposal battery please following...

Страница 4: ...Wireless Communicator 1 5m Diagram not to scale 1 5m 1 5m 2 5m Patient Environment Harmonics IEC 6100 3 2 Class A 3A m Power frequency magnetic fields should be that of a typical commercial or hospita...

Страница 5: ...BT 2 4 WS can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF Communications Equipment and the BT 2 4 BG as recommended below according to th...

Страница 6: ...d 1 2 3 Sending a Reading to SenseWear armband figure 1 figure 2 Important Information About the SenseWear biotransceiver Follow operating instructions Caution The Waste Electrical and Electronic Equi...

Страница 7: ...ommunications However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television reception which can b...

Отзывы: