14
disappears.
②
Water temperature Warning:
when the temperature of the water tank
higher than 30
℃
. Make the machine standby mode, and then use the machine after the
temperature alarm disappears.
③
Sensor alarm:
when the temperature of the water, hand piece and the
senor have malfunction, the temperature sensor will warn. Then check whether the senor
is connected properly or damaged.
④
Water flow alarm:
it will alarm when it’s wrong with the water circulation.
⑤
Hand piece alarm:
It will warn when the hand piece is not connected or
connection is failure
⑥
Gate interlock alarm:
it will warn when the interlock is open or don’t
connect the gate interlock.
⑦
Laser light warning:
it will alarm when the laser is lighting
Notice: after the exclusion of all the alarm information to the normal operation
of the laser.
8.4 Operation
8.4.1
Preparation before use
1. Before using the equipment must check the installation of the equipment to meet
the requirements of the device carefully every time.
2. Every time before using the equipment must carry the mobile phone and other
electronic products as well as the key, necklace and other metal items placed in other
places to keep properly, so as avoiding the adverse effects of the equipment operation
3. Every time you use the equipment, you must check whether the patient has a