EXT555H
TECHNIC
A
L
D
A
TA
PAGE
3
3
2
TECHNICAL DATA
2.1
General
General items
Reference data
Performance
See Table 5
Dimensions and features
See Figure 4
Maximum inlet flange temperature
80 °C
Maximum magnetic field
EXT555H DN160CF
3 mT horizontal, 5 mT vertical
EXT555H ISO160
2.5 mT horizontal, 5 mT vertical
Installation category
EN61010 part 1, category 1
Pollution degree
EN61010 part 1, category 2
Equipment type
Fixed equipment, for indoor use only
Operating altitude
up to 3000 m
Ambient temperature
5 °C - 35 °C (air cooled)
5 °C - 40 °C (water cooled)
Relative humidity
10 - 95% non condensing
Table 1 - General information
2.2
Pumping media
WARNING
Vent dangerous gases and gas
mixtures safely. Do not expose
people to these gases.
WARNING
Do not use EXT555H pumps to
pump explosive gas mixtures as the
pumps are not suitable for this
purpose.
WARNING
On an Hi variant, gas pumped
through the interstage port will mix
with gas pumped through the pump
inlet. Ensure that the gases will not
react or combine to form dangerous
gases and substances.
Note:
Concentrations of gases may be modified by
the compression of the pump.
CAUTION
Do not use an EXT555H pump to pump gases
containing more than 20% oxygen unless a bearing
purge is applied as stated in Section 2. If you do the
lubricant will be damaged and the pump will fail
prematurely.
CAUTION
Do not use the EXT555H pump to pump mercury
vapour, and do not allow mercury to come into
contact with the pump. If you do, the rotor may
corrode and fail.
Содержание EXT555H
Страница 1: ...NRA 266 880 Issue A Original 98 37 EC 89 336 EEC 73 023 EEC Instruction Manual EXT555H...
Страница 2: ......
Страница 6: ...Issue A Jun 01 EXT555H PAGE iv iv This page intentionally blank...
Страница 8: ...Issue A Jun 01 EXT555H PAGE 2 2 This page intentionally blank...
Страница 20: ...Issue A Jun 01 PAGE 14 EXT555H INSTALLATION 14 Figure 3 Maximum rate of pressure rise when venting...
Страница 24: ...Issue A Jun 01 EXT555H PAGE 18 18 This page intentionally blank...
Страница 32: ...Issue A Jun 01 EXT555H PAGE 26 26 This page intentionally blank...
Страница 40: ...Issue A Jun 01 EXT555H PAGE 34 34 This page intentionally blank...
Страница 43: ......