17
2. The height support must be locked reliably.
3. Do not adjust the stylus position using the height support during measuring. Adjust the adjustable feet to the
desired height before measuring. Measure with calipers to meet the requirements.
7.2 Height stand
Use a height stand to adjust the position between the instrument and the workpiece more accurately for more
reliable and stable operation, to expand the measuring space and measure larger workpieces. To use a height
stand, be careful as the pickup is close to the workpiece, so as to avoid damage to the pickup due to malfunction.
When the pickup is close to the workpiece surface, check the stylus position on the screen carefully. It is OK as the
pointer stays in the middle of the display. When the Ra value of the measured surface is small, it is recommended
to use the measurement platform. Start the measurement with a remote controller in order to reduce the impact of
the external environment on the measurements. To use special pickups, such as aperture, deep groove and
curved surface pickups etc., use the column height ruler height gauge or other adjustment equipment.
Height stand
7.3 Standard pickup
The standard pickup is used more, which can measure the roughness of the most flat, slope, conical surfaces,
inner holes, grooves and other surfaces. Handheld measurement is available. Except for standard pickups, the
measuring platform is required for other pickups.