11
then quit. Measure again and adjust, until the measurements are satisfied.
The calibration is subject to each range. Calibrate each range independently. The calibration results are stored in
the instrument's memory, which is not lost after shutdown.
Note: 1, To test the accompanying calibration specimen using proper measuring method, if the actual measurement
exceeds ±10% of the calibration value of the specimen, calibrate using the calibration function by the actual
error percentage with range less than ±20%.
2, In general, the instrument has been rigorously tested in the factory, and the display error is less than ± 10%. In
this case, it is recommended that users do not calibrate the tester frequently using the calibration function.
3, After the calibration value is set to “00” and confirmed, all calibration settings are clear to restore factory
settings.
3.5 Getting Started
3.5.1 Start measuring
After completion of the above steps, you can start measuring. In the main screen, press the Start button to begin
measuring.
After sampling is completed, filtering of the sampled data starts.
After filtering is completed, all the parameters are calculated.
Tips: If you accidentally touch the power button in the measurement state, the instrument will be shut down. After
booting, the pickup will be reset. Do not intervene the pickup now. After resetting, the instrument is waiting for a
new start command.
3.6 Result display
After measuring, all measurements can be observed in the following way:
3.6.1 Parameters
In the main screen, press Up to enter all Parameters Display screen. Press Up/Down to PageUp/Down. Press Set
to return to the main screen.
3.6.2 Profile Graphic
In the main screen, press Left arrow to enter the Profile Graphic screen. A sampling length is displayed on each
page. Press Left/Right arrow to switch to other sampling lengths. In this screen, press Return to Zoom In/Out the
displayed profile. Press Set/Menu button to return to the main screen.
Press Up/Down arrow to switch to the Support Curve screen. Press Set/Menu button to return to the Profile
Graphic screen.