Service Manual for RESmart
®
Auto CPAP V1.1
4/16
‘YYYY/MM/DD’=’2098/08/18’, press ‘Pressure Start/Stop Button’ when Hour Setting appears to start
the Pressure Calibration. The device output pressure is based on 20 hPa (cmH
2
O), and a number will
appear on the screen, when the output pressure is stable, if the value on the pressure meter is not 20
hPa (cmH
2
O), press ‘+/-’ User Buttons to adjust the output pressure, finally press ‘Pressure
Start/Stop Button’ to finish the Pressure Calibration.
4. Malfunction and Countermeasure
4.1 Error Code Table and Information
Error
Index
Problem Causing
Error1
Over temperature inside or Temperature sensor
fault
Malfunction on main board
Error2
Motor stop
Malfunction on motor or main board
Error3
Motor low speed
Malfunction on motor or main board
Error4
Low temperature inside
Malfunction on main board
Error7
Pressure sensor fault
Malfunction on main board
Error8
Wrong software version
Software
Error9
Host parameter error
Malfunction on main board
Error12
Over temperature inside or flowrate sensor fault
Malfunction on main board
Error13
Temperature below 0
or
℃
flowrate sensor fault
Malfunction on main board
Error14
The motor is over temperature or motor
temperature sensor fault
Malfunction on motor or
Motor temperature sensor
Error15
Motor temperature sensor fault
Malfunction on motor temperature
sensor