Service Manual for RESmart

®

 Auto CPAP V1.1 

3/16 

 

Power on the RESmart

®

 Auto CPAP device. If the device is power on from beginning, please power 

off the device and turn on again after 5 seconds more. 

4.  After the device is turned on, the following will display: 

 

Click 

Upgrade

 to start. If success, the right version number will be displayed on RESmart

®

 Auto 

CPAP device screen when every power on. 

Note: 

If failed during upgrading, please repeat from above 1-4 steps. 

If 

Error 08

 displayed after software upgrading, it means unmatched software used (for example: 

apply RESmart

®

 Auto CPAP software to RESmart

® 

CPAP device). In this case, apply correct software 

and upgrade again. 

3.2 Hidden Function 

1. 

Humidifier matching and dis-matching 

(Available on version 1.19 and later) 

Assemble humidifier on main device, without water chamber. Enter the maintenance menu, and set 
Ramp=40, then press Heated Humidifier Button when the Init P appears. Thus the humidifier is 
matched. When ‘OK’ appears on screen, humidifier matching succeeded. 

If cut the power supply after pressing ‘Pressure Start/Stop Button’ when the humidifier is matching, 
the humidifier will be dis-matched. 

2. 

Reset 

(Available on version 1.28 and later) 

Press and hold the Ramp Button until the user menu appears (about 3 seconds) when the device is 
on standby, press +/- User Buttons to access the Date Setting, set ‘YYYY/MM/DD’=’2097/07/01’, 
then press ‘Pressure Start/Stop Button’ when ‘Hour’ setting appears, thus all device settings and 
therapy records (doesn’t include Use Days and Use Time) will be reset. 

If set ‘Minute’=’16’ during the above procedure, then not only device settings and therapy records, 
but also patient information will be reset. 

If set ‘Minute’=’26’ during the above procedure, then not only device settings, therapy records and 
patient information, but also Use Days and Use Time will be reset. 

3. 

Pressure calibration

 (Available on version 1.58 and later) 

Connect pressure meter to the air outlet of the device, in the Date and Time Setting menu, set 

Содержание RESmart Auto CPAP

Страница 1: ...RESmart Auto CPAP System Service Manual...

Страница 2: ...e 2 3 1 Instruction for Upgrading Device Software 2 3 2 Hidden Function 3 4 Malfunction and Countermeasure 4 4 1 Error Code Table and Information 4 4 2 Malfunction Determination 5 Appendices 6 Appendi...

Страница 3: ...ND WARNING in this manual are to emphasize dangers to service people WARNING If do not operate properly may cause damage to people and environment CAUTION If do not operate properly may cause damage t...

Страница 4: ...l is used to help service people maintenance RESmart Auto CPAP system of BMC Medical Co more efficiently Instructions in this manual may help device work best Engineer can find needed instruction from...

Страница 5: ...following picture Please remember the software version in the red frame 1 Power off the RESmart Auto CPAP device then connect PC and the device via data cable 2 Run the software BMC RESmart Tools for...

Страница 6: ...tton when the Init P appears Thus the humidifier is matched When OK appears on screen humidifier matching succeeded If cut the power supply after pressing Pressure Start Stop Button when the humidifie...

Страница 7: ...g Error1 Over temperature inside or Temperature sensor fault Malfunction on main board Error2 Motor stop Malfunction on motor or main board Error3 Motor low speed Malfunction on motor or main board Er...

Страница 8: ...Service Manual for RESmart Auto CPAP V1 1 5 16 4 2 Malfunction Determination...

Страница 9: ...ice Fig A Main board backside Fig B Outlet leftside Hole 1 2 and 3 in Fig B connect to sensor YL and LL via silicon rubber canal Specified in Table 1 Fig B Fig A Specification of canal 1 Hole 1 on sen...

Страница 10: ...Service Manual for RESmart Auto CPAP V1 1 7 16 Fig C Power PCBA front side Sockets figuration Fig D Wire connecting on power PCBA Fig E Power supply connecting...

Страница 11: ...V1 1 8 16 Fig F Power supply to humidifier Connecting relationship Part Fig 1 2 3 4 D Back of device Fig C CN1 E Fig C CN2 Fig A P24 Fig F Left Fig F Right Blower 8 pin plug Fig A PM Table 2 2 Humidi...

Страница 12: ...Manual for RESmart Auto CPAP V1 1 9 16 Fig H Power supply to humidifier Left Right Orange Black Fig I Heater PCBA Connecting relationship Part Fig 1 2 3 4 I Fig G P2 Fig G P1 Fig H Left Fig H Right T...

Страница 13: ...Assembly the humidifier locker and button spring 4 Put fixing on the spring and fix by screw 2 3mm 8mm 5 Put power wire 6 Assembly the shield back on the bottom fix by screw 2 3mm 8mm 7 Assembly foam...

Страница 14: ...ires 14 Assembly fixing clip on the main board PCBA 15 Plug the 8 pin plug of blower in the socket PM plug 2 as in Fig E in the socket P24 on the main board PCBA 16 Assembly the main board PCBA on fix...

Страница 15: ...embly and fix the panel 23 Put on the key button 24 Assembly shield cover on the bottom adjust the position of outlet and make sure it is clamped in the right socket 25 Fix the shield by screws 4 M3 1...

Страница 16: ...ane and fix by screw 3mm 8mm Scaleboard 4 As per Table 3 assembly heater PCBA and wires 5 Put the humidifier inner connector on the shield 6 Put on light window 7 Assembly the inner and outer shield 8...

Страница 17: ...14 16 5 Humidifier Clip 1 6 Humidifier Locker 1 7 Humidifier Fixing 1 8 Filter Cover 1 9 Shield Back 1 10 Outlet 1 11 PCBA Fixing 2 12 Infrared Window 2 13 Panel 1 14 Humidifier Outer Shield 1 15 Hum...

Страница 18: ...right down 1 23 Device Pad 6 24 Blower Connector 1 25 Humidifier Inner Connector 1 26 Key Button 1 27 Humidifier Scaleboard 1 28 Heater Plane Spring 3 29 Button Spring 1 Spare Part List Part Name S N...

Страница 19: ...10010 2 PC Plastic Main Device Panel 210036 1 PC Plastic Main Device Power Supply PCBA Fixing 210015 1 PC Plastic Main Device Humidifier Outer Shield 210038 1 PC Plastic Humidifier Humidifier Inner Sh...

Отзывы: