E-20A-H-O / E-20AJ-H-O User Manual V2.1
5 / 36
Eng
lis
h
Sound Pressure Level
<
30 dB, when the device is working at the pressure of 10 hPa.
Sound Power Level
<
38 dB, when the device is working at the pressure of 10 hPa.
Maximum Flow
Test Pressures (hPa)
4
8
12
16
20
Measured Pressure at the
Patient Connection Port (hPa)
3
7
11
15
19
Average Flow at the Patient
Connection Port (L/min)
80
92
91.5
91
96
SpO
2
Range: 0
~
100%
The margin of error for SpO
2
between 70% and 100% is ± 3%. No strict accuracy requirements
for SpO
2
below 70%.
Pulse Rate
Range: 40
~
240 BPM
Margin of Error: ± 3%
Wavelengths
Red: 663 nanometers
Infrared: 890 nanometers
Maximal Optical Output Power
Less than 1.5 mw maximum average.
Tube
Length: 6 ft. (1.83 m)
The Form and the Dimensions of the Patient Connection Port
The 22 mm conical air outlet complies with ISO 5356-1.