file name and save location for this preset file. This allows for unlimited saving of presets 
under unique file names/locations. The preset function retains the channel on/off and 
channel intensity values from the save point.  

To load a preset, select “Load Preset” and select your chosen preset file from your 
specified  location.  The  “Active  Preset”  box  will  then  display  your  chosen  preset 
filename  and  preset  identification  name.  Figure  10  shows  preset  filename  as  “Test 
Presets” and preset identification name as “Test Preset ”.  

 

Figure 10: Preset Loaded 

3.13

 

Software Dual Operation Mode 

The  niji  can  receive  and  send  commands  to  multiple  sources  via  the  separate 
hardware  connection  ports  (USB  and  RS232).  For  example,  with  the  niji  software 
running on a windows desktop and a windows tablet, both devices can control and 
monitor the current level settings. If a level is changed on one software instance the 
other instance of the software will mirror the change. 

3.14

 

Using the niji with Micromanager 

The niji has a native Micromanager adapter which is available by downloading the 
nightly release. If running the niji with Micromanager using the RS232 connection, the 
baud  rate  should  be  set  to  57600  using  an  XML  configuration  file.  Please  contact 
Bluebox Optics or your local distributor for more information. 

 

Содержание niji

Страница 1: ...User Manual niji...

Страница 2: ...Driver Setup USB control only 7 3 4 Software Setup 7 3 5 Hardware Placement and Ventilation 7 3 6 Hardware Set Up 8 3 7 Connect to the niji 9 3 8 Software Control 10 3 9 Constant Optical Power Mode 1...

Страница 3: ...and String 28 8 3 Channel Level Set Example 28 8 4 Channel On Off 29 8 5 Channel Active Set Example 29 8 6 Channel Identification 29 8 7 TTL Command String 29 8 8 TTL Set Example 29 9 Technical Specif...

Страница 4: ...17 Figure 9 Preset Options 18 Figure 10 Preset Loaded 19 Figure 11 Internal Temperature Display 20 Figure 12 niji Filter Paddle 22 Figure 13 Filter Positioning 23 Figure 14 A Filter and Filter Paddle...

Страница 5: ...cation of this equipment will invalidate the warranty and may render the unit inoperable Switch off the unit when not in use Do not allow liquids to enter the housing Do not cover air vents Do not sta...

Страница 6: ...where these exist Registration WEE BU4061XA 1 4 Static Sensitivity Appropriate precautions should be taken to prevent damage from static electricity discharge 1 5 Technical Support For technical supp...

Страница 7: ...ision Light output can be coupled into a liquid light guide LLG The niji system also comprises a TTL port whereby a TTL signal can trigger light output switching on off The niji multi wavelength light...

Страница 8: ...plication allows the user to control each individual channel software is provided with each unit and written exclusively for Bluebox Optics 3 3 Driver Setup USB control only The drivers for this unit...

Страница 9: ...al temperature probes are installed to monitor the equipment temperature and this data is displayed via the software interface 3 6 Hardware Set Up Connect the liquid light guide LLG to the device via...

Страница 10: ...several built in safety conditions which must be satisfied before any light is emitted When the power indicator at the front of the unit flashes red the following actions should be taken Ensure that...

Страница 11: ...s they are in Figure 4 When a single or multiple channels are switched on the status will change to LIGHTS ON This is an integral safety feature by design See Table 1 for all available light source co...

Страница 12: ...vely Figure 4 Lights OFF All Channels OFF Figure 5 Lights ON All Channels On Seven channels are available Channel 5 delivers highly broadband light and as standard is configured in the broadest band a...

Страница 13: ...m to low optical power applciations This can be enabled in the Configuration window In Constant Optical Power mode the niji can be controlled using the sliders etc as it would normally be used but the...

Страница 14: ...13 www BlueBoxOptics com V2 4 niji...

Страница 15: ...programmable with many microscope systems Custom configuration including intial power on settings can be achieved using the XML configuration file If you wish to use the niji in compatible mode or wi...

Страница 16: ...off when 5V is applied The level of emission of each channel is set either through the niji software or through serial commands Those channels that are not connected to the TTL interface will remain...

Страница 17: ...the TTL triggered channels will be controlled by the TTL interface In this mode only those channels which are tied to 5V are in the off state Active High Active Low 100k Pull Up 100k Pull Down 5V GND...

Страница 18: ...17 www BlueBoxOptics com V2 4 niji Figure 8 TTL Mode Active...

Страница 19: ...be saved and recalled at any time however an unlimited number of preset files can be saved if needed see Figure 91 Figure 9 Preset Options To save a preset select your chosen preset identification na...

Страница 20: ...reset Loaded 3 13 Software Dual Operation Mode The niji can receive and send commands to multiple sources via the separate hardware connection ports USB and RS232 For example with the niji software ru...

Страница 21: ...ternal Temperature The internal temperature of the unit is displayed in the information tab Case Temperature 2 monitors ambient case temperature Case Temperature 1 monitors the temperature at board le...

Страница 22: ...level where 100 represents the maximum available intensity and 1 represents the minimum emission of the system 4 3 Light Guide Adapter and Interlock To install the light guide push the 3mm light guid...

Страница 23: ...dle shaped material which is safe to handle Figure 12 Figure 12 niji Filter Paddle One side of the paddle is rough while the other is smooth The smooth surface should face towards the module with the...

Страница 24: ...w and set to one side Gently release the filter by pushing the filter from the smooth to the rough side of the paddle Do not make contact with the surface of the filter itself but instead apply force...

Страница 25: ...gure 15 Place the paddle on a smooth and clean surface remove the screw and set to one side Gently release the filter by pushing through from the smooth side of the paddle Before inserting a filter po...

Страница 26: ...replace the lid as soon as possible Each module is labelled with its channel number and name indicating its location in the niji This labelled area also acts as a grip for the removal and insertion o...

Страница 27: ...made via a card edge connection this will need to be gently eased out Do not allow thumb screws to roll into the unit from the top plate Set them to the side once removed Store any unused modules in...

Страница 28: ...cleaning procedure should the surface of the filters be contaminated through handling exposure to dust etc Full instructions can be found on our website www blueboxoptics com 8 Serial Command Control...

Страница 29: ...port should be configured as follows Port select COM Baud rate 115200 Data bits 8 Stop Bit 1 Parity None Flow control RTS CTS Forward None Transmitted CR and LF Figure 19 Port Setup 8 2 Channel Level...

Страница 30: ...avelength 1 Ultra Violet 2 Royal Blue 3 Blue 4 Cyan 5 Green Amber 6 Red 7 Far Red 8 7 TTL Command String The TTL command string is split into four parts These are the command identifier function ON OF...

Страница 31: ...Mercury Free Yes Operating Conditions 10 to 40 C Storage Temperature 10 to 60 C Humidity 40 to 70 RH Warranty 12 Months 3 year Optional 9 1 External Power Supply Manufacturer Bluebox Optics Ltd Model...