4.3.9  

Proportional Dosing

Used to configure proportional Parts Per Million dosing to a system. This method 
of proportional dosing is based off an input signal the pump is receiving from an 
external flow meter. The flow meter must have a high speed pulse output >10Hz. 
You will need to refer to flow meter instruction manual to obtain the K-factor for 
the flow meter.

Default settings:  

12.5% dose solution concentration

 

 

 

 

 

1.25 dose solution Specific Gravity

 

 

 

 

 

5.0 K-factor (Pulses Per Liter), 

see flow meter instruction manual

 

 

 

 

 

1.0 Parts Per Million to dose

Press MODE button until 

Proportional Dosing 

is displayed in the top line of the 

display. 

To configure the pump, navigate to 

Proportional Dosing 

menu by using the 

short-cut method described at the beginning of the section, or by pressing 
MENU button, then selecting Input Setup, Input Modes, and then 

Proportional

.

Press UP or DOWN arrow to scroll through 0 - 9 on selected digit.

Press RIGHT arrow to scroll over to next digit to the right. If you pass your 
desired digit, you can easily scroll back by continuously pressing RIGHT button.

Press ENTER to save the changes.

Continue this process until all four screens have been configured.

If you used the short-cut to enter Proportional Dosing setup, press and hold 
Mode button until Run Mode screen is displayed.

If you used the Menu button to navigate to Proportional Dosing setup, you must 
navigate back out of menu structure. To do this you must select <-Esc at bottom 
of every screen menu until you see Run Mode screen displayed.

- Proportional Dosing

Input K-Factor (pulses):

Pulses Per Liter.
0.0001  -  9999.9999

- Proportional Dosing

Dose Specific Gravity:

Specific Gravity of solution.
Press ENTER to set.

.25

1

- Proportional Dosing

Dose Concentration:

Percentage of concentration.

Press ENTER to set.

- Proportional Dosing

PPM to dose:

Parts Per Million to dose.
000.0  -  100.0 PPM

0

12.5%

0

005.0000

0

01.0

Injection
Point

S3 Series  |  Hybrid Ultrasonic Flowmeter

NEMA 4X

IP 66

Industries, Ltd.

ProSeries

by Blue-White Ind.

TM

Doppler /
Transit Time

Discharge

Line

Paddlewheel

Flow Sensor / Meter

Ultrasonic

Flow Meter

Injection
Point

Example #1, Ultrasonic flow meter

Example #2, Paddlewheel flow meter

Discharge

Line

or

- Menu

+ Configuration

+ Input Setup

+ Pump Output

<- Esc

- Input Setup

+ Contact Closure Input

- Contact Closure Input

    Close: Stop Pump
    Open: Stop Pump

<- Esc

Hidden

+ Motor - Select RPM
+ Tubing - Max Flowrate

+ Set FVS

+ Operating Input Modes

    Disable

Done

4.4   Contact Closure Input (Remote Start/Stop)

Used to remotely start and stop the pump using a close=stop or open=stop 
signal. If the pump must start when the loop is open, then select “Close: Stop 
Pump” option. Can be used with an external foot pedal, a PLC, contact closure, 
or other similar external devices.

Default settings:  

Disable

CC Input Range:  

6 - 30 VDC

 

 

 

 

 

or

 

 

 

 

 

Dry Contact Closure (no voltage required)

 

 

 

 

 

[See section 5.1 for wire connections]

Navigate to 

Contact Input 

menu by MENU button, then selecting Input Setup, 

and then 

Contact Input

.

Press UP or DOWN arrow to scroll through your options.

Press ENTER to make a selection. You should then notice the radio button 
(square box) is now filled in next to your selection.

Press DOWN arrow to scroll down to Done selection. Then press ENTER.

To prevent false triggering due to flickering (high speed) electrical switches, a 
trigger delay time can be configured to delay the pump command. The delay 
time unit of measure is seconds. A two second delay time is recommended.

To navigate back out of menus, select <-Esc and press the ENTER button at the 
bottom of every screen menu until you see the Run Mode screen displayed.

IMPORTANT: To begin operation, press the START button to place pump in 
STANDBY. The display background will turn blue indicating the pump has been 
stopped remotely. When the pump is started by the remote contact, the display 
background will turn green.

IMPORTANT: If the Contact Closure Input is enabled, the pump will display 
STANDBY if the pump has been stopped by the Remote Contact Closure. 

Please use caution in this mode as the pump may Start at anytime. If you 
must perform maintenance to the pump, Press STOP button.

When Contact Closure Input is enabled, the word 

Remote

 will always be 

displayed on the lower left side of the display screen.

Manual Speed Adjust

Waiting for SIGNAL...

STANDBY

REMOTE

Manual Speed Adjust

GAL/DAY

228

Å

É

ÄÈ

REMOTE

Signal stopped pump

Signal started pump

Remote Delay

Input Remote Delay

Max Delay = 999.9 (s)
Press ENTER to set

0

00.0

Page 24

Page 25

®

CHEM-FEED  MD3

®

CHEM-FEED  MD3

Содержание CHEM-FEED MD3 Series

Страница 1: ...Diaphragm Metering Pump Series MD3 CHEM FEED...

Страница 2: ...3 1 Manual Adjust manual speed adjust 16 4 3 2 4 20 mA Input 17 4 3 3 0 10 VDC Input 18 4 3 4 Frequency Input Hz 19 4 3 5 Pulse Batch low speed batch 20 4 3 6 Manual Cycle Adjust repeating cycle time...

Страница 3: ...m Model Number Matrix READ THE INSTRUCTION MANUAL PRIOR TO INSTALLATION AND USE 1 714 893 8529 sales blue white com customerservice blue white com 5300 Business Drive Huntington Beach CA 92649 Page 3...

Страница 4: ...115V60Hz NEMA 5 15 USA 230V60Hz NEMA 6 15 USA 220V50Hz CEE 7 VII EU 240V50Hz AS 3112 Australia New Zealand Pump Head Assembly Pump Head PVDF Manifold PVDF Adapter Connections PVDF Valve Cartridges PVD...

Страница 5: ...ite s output Flow Verification Sensor FVS system Includes stainless steel extended mounting brackets Lifts pump 4 1 2 11 43 cm for easy access in hard to reach areas Enclosure Rating NEMA 4X Construct...

Страница 6: ...n check valve at discharge side of pump to prevent system fluid from flowing back through pump during pump maintenance A pressure relief valve is recommended at discharge of pump Bolt pump to a stable...

Страница 7: ...0cm 4 00in 10 15cm n Made out of tough Stainless Steel n Provides a stable mounting surface Stainless Steel extended brackets allow the pump to be securely mounted to most any surface floor shelf or s...

Страница 8: ...d and sealed Never strap control input output cables and power cables together Power Interruption This pump has a user programmable auto restart feature which will either restore the pump to the opera...

Страница 9: ...AMP MAX AT 125 VOLT AC 0 8 AMP MAX AT 30 VOLT DC 5A SLOW BLOW 20 X 5MM 1 2 3 T1 PULSE GND 4 20 1 2 3 T1 PULSE GND 4 20 0 10 15VDC 1 2 T2 0 1000 HZ MAX NEGATIVE POSITIVE POSITIVE NEGATIVE NO VOLTAGE 96...

Страница 10: ...elect menu items increase menu values by one and increase pump output in Manual Speed Adjust mode Press and hold to increase values progressively faster Press and release to stop the pump Press and re...

Страница 11: ...e While making a selection where only one choice is allowed you will see a radio button Radio button example In a screen where you are making changes you will see the word Done located at the bottom o...

Страница 12: ...e menu structure and enter the run mode Configuration Display Language Set Password Display Rate Esc Reset to Default Display Language English Spanish German Done French Display Language Configuration...

Страница 13: ...gled through Speed RPM and your selected Display Rate by pressing the right arrow button Press MENU button to enter the menu structure Select Configuration and Press ENTER button Select Display Rate a...

Страница 14: ...wrate Flowrate ML min Press ENTER to set 0000 0 1 Contact Input Close Stop Pump Open Stop Pump Input Time Delay OFF 000 Range 1 255 sec Disable 00 0 Input Setup 4 20 mA Input Frequency Input Pulse Inp...

Страница 15: ...rate Input Modes Contact Input Set FVS Esc Hidden Remote Local Set DFD Set Rev Alarm Max Flowrate Input Max Flowrate Flowrate ML min Press ENTER to set 0000 0 1 To calculate the Max Flowrate To determ...

Страница 16: ...ack out of the menu structure To do this you must select Esc at the bottom of every screen menu until you see the Run Mode screen displayed Tip The Manual Speed Adjust mode can be combined with Contac...

Страница 17: ...process until all four screens have been configured If you used short cut to enter 4 20 mA input setup press and hold the Mode button until the Run Mode screen is displayed If you used Menu button to...

Страница 18: ...top line of the display To configure the pump navigate to 0 10 VDC Input menu by using the short cut method described at the beginning of the section or by pressing MENU button then selecting Input Se...

Страница 19: ...select the digit to change press UP or DOWN arrow to scroll through 0 9 on selected digit Press enter to save changes and move to the next input screen To input flow rate values press the RIGHT ARROW...

Страница 20: ...k by continuously pressing the RIGHT button Press ENTER to save the changes Continue this process until all three screens have been configured If you used the short cut to enter Pulse Batch setup then...

Страница 21: ...onfigured If you used the short cut to enter Manual Cycle Adjustment setup press and hold Mode button until Run Mode screen is displayed If you used the Menu button to navigate to Cycle Adjustment set...

Страница 22: ...xt digit to the right If you pass your desired digit you can easily scroll back by continuously pressing RIGHT button Press ENTER to save the changes Continue this process until two screens have been...

Страница 23: ...arrow to scroll over to next digit to the right If you pass your desired digit you can easily scroll back by continuously pressing RIGHT button Press ENTER to save the changes Continue this process un...

Страница 24: ...esired digit you can easily scroll back by continuously pressing RIGHT button Press ENTER to save the changes Continue this process until all four screens have been configured If you used the short cu...

Страница 25: ...gering due to flickering high speed electrical switches a trigger delay time can be configured to delay the pump command The delay time unit of measure is seconds A two second delay time is recommende...

Страница 26: ...n discharge side of MD3 pump Wiring for sensor can be connected directly to a MD3 pump Pump will stop pumping if sensor detects no flow A relay will then close allowing for remote alarm indication or...

Страница 27: ...an 430 microsiemens will not be detected This system is capable of detecting presence of a large number of chemicals including Sodium Hypochlorite Chlorine Hydrochloric muriatic Acid Sodium Hydroxide...

Страница 28: ...etting section 4 4 is switched to disabled while LOCAL ONLY is selected This menu setting will return to the previous setting when REMOTE ONLY or DISABLED is selected Input Setup Input Modes Contact I...

Страница 29: ...Stop Frequency Mode Cycle Timer Mode Dispensing Mode Manual Dosing Prop Dosing 0 10 VDC Mode Pulse Mode Disable FVS DFD Pump Run Stop 4 20mA Mode Manual Mode Frequency Mode Cycle Timer Mode Dispensing...

Страница 30: ...0 Pump Output 4 mA 100 Pump Output 20 mA Signal Output Frequency Output Disable Done 4 20 mA Output Default 4 20 mA output settings shown Signal Output Disable Done 4 20 mA Output Default Frequency H...

Страница 31: ...r does not assume responsibility for damage to pump that has been caused by chemical attack 6 2 Cleaning Pump Pump will require occasional cleaning especially Injection fitting Footvalve Strainer and...

Страница 32: ...7 0 Exploded view and parts list 1 2 4 6 8 7 10 22 9 15 14 3 13 11 3 12 Page 32 CHEM FEED MD3...

Страница 33: ...001 505 MANIFOLD WITH FLANGE AND UNION NUT PLUG WITH O RING EP 6 90002 350 DRIVE HEAD INSERT 2 7 72000 583 KIT DIAPHRAGM PVDF MD3 2 8 70001 345 CARTRIDGE VALVE TFE P 8 70001 346 CARTRIDGE VALVE EP 9 9...

Страница 34: ...32 151 303 751 1508 2252 3041 RPM 1 3 10 19 48 95 142 190 GPH 0 5 2 4 4 8 11 9 23 9 35 7 48 2 Feed Rate at 90 PSIg ML MIN 25 145 278 694 1388 2107 2391 RPM 1 3 10 19 48 95 142 190 GPH 0 4 2 3 4 4 11 0...

Страница 35: ...Page intentionally left blank Page 35 CHEM FEED MD3...

Страница 36: ...Page intentionally left blank Page 36 CHEM FEED MD3...

Страница 37: ...Page intentionally left blank Page 37 CHEM FEED MD3...

Страница 38: ...iption of the problem and proof of purchase Prepay all shipping and insurance cost COD shipments will not be accepted Damage caused by improper packaging is the responsibility of the sender When In Wa...

Страница 39: ...ia New Zealand 9 230V 50HZ power cord BS 1363 A plug United Kingdom X No Power Cord Elastomer Material O Rings V TFE P E EP Miscellaneous Options leave blank for standard model A Hastelloy C check val...

Страница 40: ...ramework available to them for the return recycle recovery of WEEE and minimize any potential effects of EEE on the environment and human health due to the presence of hazardous substances The WEEE ma...

Отзывы: