6.5 Mode 2 - 4-20 mA input -
In this mode, the on-time of the cycle will
automatically adjust to match the received mA input value. When the mA
input value is equal to the programmed maximum, the pump will run
continuously.
Four values must be programmed:
1)
ON-T
= The amount of time the pump will run, per cycle, when the
minimum mA value is received. (Typically programmed to zero)
2)
mA minimum
= The mA input value that will result in the on time (
ON-
T
). (Typically programmed to 4 mA)
3)
TOT-T
= The total cycle time.
4)
mA maximum
= The mA input value that will result in the pump running
continuously.
Example:
ON-T setting = 0 seconds
mA minimum setting = 4mA
TOT-T setting = 8 seconds
mA maximum setting = 14.8mA
6
4
8 10 12 14 16 18 20
0
2
4
6
8
Pump Run-T
ime (Sec.)
Milliamp input (mA)
Page 14
Page 15
+
Set the pump for mode 2
. Press the
MODE button until
MODE 2
is shown on
the LCD display.
+
Enter the programming mode.
At the
same time, press the RUN/STANDBY
button and the MODE button. A blinking
ARROW
will point to the word
PROGRAM
indicating the program mode has been
activated. The on time
ON-T
icon will
blink. The currently selected time unit
+
Press the MODE button to exit the on time
programming screen and enter the mA
minimum programming screen. The
mA
icon will blink. A blinking
ARROW
will
appear next to the word
MINIMUM
. The
current minimum mA setting will be
displayed and the left most (selected) digit
will blink.
RUN
PROGRAM
STAND-BY
PRIME
MINIMUM
MAXIMUM
2
00.0
MODE
ON-T
SEC
icon will be displayed. The current on time setting will be displayed and
the left most (selected) digit will blink. Note: The left most digits can be
programmed from 0 - 19. The decimal is fixed and cannot be moved.
+
Pressing the DIGIT button will increase the selected digit.
+
Pressing the FIELD button will select a new digit to the right or the time
unit.
+
Press the DIGIT button to increase the selected digit or time unit.
RUN
PROGRAM
STAND-BY
PRIME
MINIMUM
MAXIMUM
2
04.0
mA
MODE
+
Pressing the DIGIT button will increase the selected digit.
+
Pressing the FIELD button will select a new digit to the right.
+
Press the DIGIT button to increase the selected digit.
(selected) digit will blink. Note: The left most digit can be programmed
from 0 - 9. The decimal is fixed and cannot be moved.
+
Pressing the DIGIT button will increase the selected digit.
+
Pressing the FIELD button will select a new digit to the right or the time
unit.
+
Press the DIGIT button to increase the selected digit or time unit.
+
Press the MODE button to exit the mA
minimum programming screen and enter
the total time programming screen. The
total time
TOT-T
icon will blink. The
currently selected time unit icon will be
displayed. The current total time setting
will be displayed and the left most
RUN
PROGRAM
STAND-BY
PRIME
MINIMUM
MAXIMUM
2
00.0
MODE
SEC
TOT-T
RUN
PROGRAM
STAND-BY
PRIME
MINIMUM
MAXIMUM
2
20.0
mA
MODE
+
Press the MODE button to exit the total
time programming screen and enter the
mA maximum programming screen. The
mA
icon will blink. A blinking
ARROW
will appear next to the word
MAXIMUM
.
The current maximum mA setting will be
displayed and the left most (selected) digit
will blink.
+
Pressing the DIGIT button will increase the selected digit.
+
Pressing the FIELD button will select a new digit to the right.
+
Press the DIGIT button to increase the selected digit.
+
At the same time, press the RUN/STANDBY button and the MODE
button. A blinking
ARROW
will point to the word
RUN
indicating the run
mode has been activated.
ø
NOTE:
If while in the program mode no buttons are pressed within 60
seconds, the circuitry will automatically return to the run mode.
C-1100E