Enable and Program the FVS System
The FVS system must be enabled and an alarm delay time programmed.
+
Set the pump for mode 0
. Press the MODE button until
MODE 0
is
shown on the LCD display.
+
Enter the programming mode.
At the same time, press the
RUN/STANDBY button and the MODE button. A blinking
ARROW
will
point to the word
PROGRAM
indicating the program mode has been
activated.
C-1100E
Page 12
C-1100E
Page 13
decrease the number of seconds. The number will increase to a maximum
of 256 seconds and roll over to OFF. To turn off the system, set the
number to 000 or OFF.
+
To exit the programming mode, press the RUN/STANDBY button and the
MODE button at the same time. The arrow next to the word
PROGRAM
will disappear and an arrow will appear next to the word
RUN
.
ø
NOTE:
If while in the program mode no buttons are pressed within 10
seconds, the circuitry will automatically return to the run mode.
+
The TFD system .
The
TFD
icon will blink.
The TFD system is only used on A-100N
series peristaltic pumps. The word
ON
will
display indicating the TFD system is
activated. No changes are necessary.
Although it is not necessary, you may press
the DIGIT button to toggle the TFD system
on and off.
+
The FVS system .
Press the MODE button
to enter the FVS system programming. The
FVS
icon will blink. The display will
indicate the current alarm delay time setting
in seconds. (000 = OFF).
+
Press and hold he DIGIT button to increase
the number of seconds of alarm delay time,
press and hold the FIELD button to
RUN
PROGRAM
STAND-BY
PRIME
MINIMUM
MAXIMUM
0
On
MODE
TFD
RUN
PROGRAM
STAND-BY
PRIME
MINIMUM
MAXIMUM
0
001
MODE
FVS
SEC
6.4 Mode 1 - Manually adjusting the output
-
In this mode, the pump
is turned on and off by an electronic cycle timer. The pump will energize for
the duration of the “on time” and de-energize for the remainder of the “total
time” thus completing one cycle. The cycle then repeats.
The “on time” and “total time” cycles are independently adjustable from 0.1
to 199.9 units of measure with a 0.1 unit resolution. The units of measure
can be seconds, minutes, hours or days.
Example: If the “total time” cycle is adjusted for 90 seconds and the “on
time” portion of the cycle is adjusted for 5 seconds, the pump will run for 5
seconds and turn off for 85 seconds (90 second total cycle). This cycle is
repeated until either the standby button is pressed, the cycle time is changed
or the input power is disconnected from the pump.
+
Set the pump for mode 1
. Press the MODE button until
MODE 1
is
shown on the LCD display.
+
Enter the programming mode.
At the same
time, press the RUN/STANDBY button and
the MODE button. A blinking
ARROW
will
point to the word
PROGRAM
indicating the
program mode has been activated. The total
time
TOT-T
icon will blink. The currently
RUN
PROGRAM
STAND-BY
PRIME
MINIMUM
MAXIMUM
1
199.9
MODE
TOT-T SEC
RUN
PROGRAM
STAND-BY
PRIME
MINIMUM
MAXIMUM
1
42.5
MODE
ON-T
SEC
selected time unit icon will be displayed. The current total time setting will
be displayed and the left most (selected) digit will blink. Note: The left most
digit can be programmed from 0 - 19. The decimal is fixed and cannot be
moved.
+
Pressing the DIGIT button will increase the selected digit.
+
Pressing the FIELD button will select a new digit to the right or the time
unit.
+
Press the DIGIT button to increase the selected digit or time unit.
+
Press the MODE button to exit the total
time programming screen and enter the on
time programming screen. The
ON-T
icon
will blink. The currently selected time unit
icon will be displayed. The current on time
setting will be displayed and the left most
(selected) digit will blink.
+
Pressing the DIGIT button will increase the selected digit.
+
Pressing the FIELD button will select a new digit to the right or the time
unit.
+
Press the DIGIT button to increase the selected digit or time unit.
+
At the same time, press the RUN/STANDBY button and the MODE
button. A blinking
ARROW
will point to the word
RUN
indicating the run
mode has been activated.
ø
NOTE:
If while in the program mode no buttons are pressed within 60
seconds, the circuitry will automatically return to the run mode.
Contact Closure Alarm Output -
A contact closure output (3A 125VAC
relay) is provided with the FVS system. The relay can be configured for
normally open (factory default) or normally closed operation by properly
positioning the connector plug on the circuit board (see page 7).