Blue-White APFCL Скачать руководство пользователя страница 7

2

Engineering Specifications

Item

APFCL

Sensor Body Material

304SS

Sensor Name

FCL700

Chlorine Range

0.00-5.00 ppm, 0.00-10.0 ppm (Free or Total)

Precision

+/- 0.01mg/L or 1% of the value w/pH compensation to 9.0

pH Range

0-14

pH Precision

+/- 

0.01 pH

Sample Operating Temperature

40 – 104 °F  (4 °C – 40 °C)

Sample Inlet Pressure

7.25 – 30 psi  (0.05 – 0.2MPa)

Sensor Maximum Pressure

100 psi  (6.9 Bar)

Sensor Response Time

T95≤60s – Free Chlorine / T95≤5s - pH

Measurement Interval

Continuous

Installation

WR05 Self-Regulating Flow Reservoir 

w/Rotameter & Isolation Valve - Included

WR05 Minimum Flow Rate

3.1 g/h  (200 mL/minute)

WR05 Maximum Flow Rate

6.2 g/h   (400 mL/minute)

WR05 Sample Inlet

1/4 - inch OD

WR05 Sample Outlet

3/4 - inch ID - To Drain

WR05

 Sewage Overflow Outlet

3/8 - inch OD - To Drain

Panel Power Supply

110/220VAC  50/60 Hz, 0.6A

Panel Storage Temperature

-4 – 158 

o

F (-20 – 70 

o

C)

Panel Operating Temperature

32 – 122 

o

F (-0 – 50 

o

C)

TS05 Display

2.8” Color 320 x 240 Resolution

Input

2 x 4-20 mA / RS-485 Modbus-RTU

Output

2 x 4-20 mA (Cl and pH)/ RS-485 Modbus-RTU / 1 x Contact Relay

Data Storage

32 M Flash

USB

1 x USB host for data downloading

Relative Humidity

5% - 95% (No Condensation)

Altitude

<6,561 feet (<2,000 Meter)

Dimension (H x W x D)

Panel 24.0” x 12.0” x 15.67” (610H x 305W x 398D mm)

Approximate Weight

 ~ 11 lbs (5 kg)

Wet Material

UPVC / Polycarbonate

Rating

IP-65 Panel-Display / IP-67 Sensors

Compliance

EPA 334.0 / ISO 7393

Regulation

CE Marked / RoHS

Typical Electrode Service Life

2 years

Electrode Warranty

6 Months

Sensor Body Warranty

13 Months

Shipping Dimensions

29” x 17” x 13” (737H x 432W x 330D mm)

Engineering Specifications

Содержание APFCL

Страница 1: ...Quick Start Guide APFCL Free Chlorine Turbidity Analyzer Panel...

Страница 2: ...ate real time measurement display and data logging of Free Chlorine pH and Temperature utilizing proprietary smart sensor technology coupled with a touch screen display and data log ging terminal Refe...

Страница 3: ...rs Electrical Schematics Quick Start Guide Tools Needed Follow these instructions before installing your analyzer to avoid failure Read this guide and Instruction Manual before using CAUTION The APFCL...

Страница 4: ...ow Connect the inch weir overflow tubing Outlet Line Connect 3 4 ID tubing to the outlet drain This is the sample water outlet flow This line must be diverted to drain Wall Mount Space The APFCL analy...

Страница 5: ...les in concrete and clean thoroughly Located strut and mount with appropriate tools Check to make sure strut is secure and level 2 4 Connect Electrical and Tubing Use provided electrical diagram to co...

Страница 6: ...ptions press the BACK Key and refer to Instruction Manual for more informa tion regarding features and user management NOTE The FCL Sensor has a 5 minute Initialization period before readings begin Se...

Страница 7: ...nch ID To Drain WR05 Sewage Overflow Outlet 3 8 inch OD To Drain Panel Power Supply 110 220VAC 50 60 Hz 0 6A Panel Storage Temperature 4 158 o F 20 70 o C Panel Operating Temperature 32 122 o F 0 50 o...

Страница 8: ...ssive heat or pressure or allowing the unit to support the weight of related piping BLUE WHITE does not assume responsibility for any loss damage or expense directly or indirectly related to or arisin...

Отзывы: