Installation Requirements

Power Supply:

 100~240VAC 50/60Hz

Water Supply: 

Inlet water pressure should be from 7.25 – 30 psi (0.05-0.2MPa) 

with an inlet feedwater line diameter of ¼-inch O.D. Tubing. The APFCL is pro-

vided with an inlet Rotameter and flow regulating valve for sample water inlet 

flow control and limited pressure regulation. The range of inlet flow for the WR05 

should be consistently maintained between 200 and 400 mL per minute.

Inlet Water:

 Connect the ¼-inch inlet water tubing to the quick adapter provided. 

Consistent flow of 200-400mL is required.

Drainage:

 The WR05 outlet tube (3/4” ID Tubing) located on the bottom of the 

WR05, as well as the WR05 weir overflow (3/8-inch O.D. Tubing) located on the 

bottom of the WR05, should both be connected to a discharge drain via gravity 

flow.

Weir Overflow: Connect the ⅜-inch weir overflow tubing.

Outlet Line:

 

Connect 3/4” ID tubing to the outlet drain. This is the sample water 

outlet flow. 

This line must be diverted to drain.

Wall Mount Space: 

The APFCL analyzer panel size is roughly 12” H x 24” W 

x 10” D  in dimension. Please accommodate sufficient space for mounting. The 

panel is equipped with pre-installed rear unistrut for simple wall mounting.

Wall Mount Weight:

 Approximately 45 lbs (20kg). Please use appropriate 

mounting hardware.

Wiring

The APFCL analyzer has universal AC power supply equipment allowing 

users simply to plug the power supply into a 100~240V AC 50/60Hz power 

outlet for normal operation.

The process of electrical connection to contact the 220V single-phase power 

supply, should be operated by personnel with an electrician's license. Failure 

to operate according to the electrical code of practice may result in electric 

shock injury or even death. 

Note: When in doubt regarding your electrical 

installation, contact a licensed electrician.

Содержание APFCL

Страница 1: ...Quick Start Guide APFCL Free Chlorine Turbidity Analyzer Panel...

Страница 2: ...ate real time measurement display and data logging of Free Chlorine pH and Temperature utilizing proprietary smart sensor technology coupled with a touch screen display and data log ging terminal Refe...

Страница 3: ...rs Electrical Schematics Quick Start Guide Tools Needed Follow these instructions before installing your analyzer to avoid failure Read this guide and Instruction Manual before using CAUTION The APFCL...

Страница 4: ...ow Connect the inch weir overflow tubing Outlet Line Connect 3 4 ID tubing to the outlet drain This is the sample water outlet flow This line must be diverted to drain Wall Mount Space The APFCL analy...

Страница 5: ...les in concrete and clean thoroughly Located strut and mount with appropriate tools Check to make sure strut is secure and level 2 4 Connect Electrical and Tubing Use provided electrical diagram to co...

Страница 6: ...ptions press the BACK Key and refer to Instruction Manual for more informa tion regarding features and user management NOTE The FCL Sensor has a 5 minute Initialization period before readings begin Se...

Страница 7: ...nch ID To Drain WR05 Sewage Overflow Outlet 3 8 inch OD To Drain Panel Power Supply 110 220VAC 50 60 Hz 0 6A Panel Storage Temperature 4 158 o F 20 70 o C Panel Operating Temperature 32 122 o F 0 50 o...

Страница 8: ...ssive heat or pressure or allowing the unit to support the weight of related piping BLUE WHITE does not assume responsibility for any loss damage or expense directly or indirectly related to or arisin...

Отзывы: