Start-Up and Operation

After the TS-05 is turned on, the main interface will display the chlorine residu-

al, pH value, and temperature of the solution. The green dot indicates that the 

communication between the TS-05 and the probe is normal.

 To access other 

options, press the BACK Key and refer to Instruction Manual for more informa-

tion regarding features  and user management.

NOTE:

 The FCL Sensor has a 5 minute “Initialization period” before readings 

begin. 

Settings

Upon pushing the Back Key, the Settings options will appear. For further pro-

gramming instructions and a functional overview of each section, please refer to 

complete Instruction Manual. 

Main Key

Up Key

Back Key

Enter Key

Down Key

Real Time Monitoring

Содержание APFCL

Страница 1: ...Quick Start Guide APFCL Free Chlorine Turbidity Analyzer Panel...

Страница 2: ...ate real time measurement display and data logging of Free Chlorine pH and Temperature utilizing proprietary smart sensor technology coupled with a touch screen display and data log ging terminal Refe...

Страница 3: ...rs Electrical Schematics Quick Start Guide Tools Needed Follow these instructions before installing your analyzer to avoid failure Read this guide and Instruction Manual before using CAUTION The APFCL...

Страница 4: ...ow Connect the inch weir overflow tubing Outlet Line Connect 3 4 ID tubing to the outlet drain This is the sample water outlet flow This line must be diverted to drain Wall Mount Space The APFCL analy...

Страница 5: ...les in concrete and clean thoroughly Located strut and mount with appropriate tools Check to make sure strut is secure and level 2 4 Connect Electrical and Tubing Use provided electrical diagram to co...

Страница 6: ...ptions press the BACK Key and refer to Instruction Manual for more informa tion regarding features and user management NOTE The FCL Sensor has a 5 minute Initialization period before readings begin Se...

Страница 7: ...nch ID To Drain WR05 Sewage Overflow Outlet 3 8 inch OD To Drain Panel Power Supply 110 220VAC 50 60 Hz 0 6A Panel Storage Temperature 4 158 o F 20 70 o C Panel Operating Temperature 32 122 o F 0 50 o...

Страница 8: ...ssive heat or pressure or allowing the unit to support the weight of related piping BLUE WHITE does not assume responsibility for any loss damage or expense directly or indirectly related to or arisin...

Отзывы: