Blue-White industries flexflo A-100NV Скачать руководство пользователя страница 16

A-100N

7.4.1

How to Remove the Old Pump Tube

The pump roller assembly spins in a counter clockwise direction.  The pump head inlet (suction) side is 
located at the bottom of the pump and the outlet (discharge) is located at the top of the pump head.

7.4.1.1

Release any pressure that may be in the 
discharge tubing.

7.4.1.2

Disconnect the suction and discharge tubes 
from the pump tube.

7.4.1.3

Remove the pump head cover.

7.4.1.4

With the pump running, pull the inlet fitting 
out of the pumphead. Guide the tube counter 
clockwise away from the rollers. Pull the 
outlet fitting out of the pump head.

7.4.2

How to Install a New Pump Tube

 

Be sure the pump head chamber is clean and free of any debris.  Remove and inspect the roller assem-
bly.  Be sure the rollers spin freely.  If required, apply a small amount of grease to the pump head cover 
bearing.  Silicone oil 

only 

may be used on the tubing and rollers. See section 7.2.

6.4.2.1

With the pump running, insert the inlet (suction) side of the Pump Tube fitting into the lower retaining 
slot in the pump head. Fig. 7.2.

6.4.2.2 Carefully

 guide the Pump Tube into the pump head.  Stretch the tube slightly and insert the outlet 

(discharge) fitting into the upper retaining slot in the pump head.  Fig. 7.3.

6.4.2.3

Place the clear cover on the pump head and secure with three screws.

FIG. 7.2

FIG. 7.3

Page 17

7.4

How to Replace the Pump Tube 

The pump tube assembly will eventually break if not replaced.  The tube has been designed for a 
minimum service life of 500 hours.  However, the life of the tube is affected by many factors such as the 
type of chemical being pumped, the amount of back pressure, the motor RPM, temperature and others.  
The pump tube assembly must be inspected and replaced regularly.

After replacing the pump tube, press the RUN/STANDBY button and the DIGIT button at the 
same time to reset the service warning timer.

A-100N

7.0

How to Maintain the A-100NV

7.1

Routine Inspection and Maintenance

The A-100NV requires very little maintenance.  However, the pump and all accessories should be 
checked weekly.  This is especially important when pumping chemicals.  Inspect all components for 
signs of leaking, swelling, cracking, discoloration or corrosion.  Replace worn or damaged components 
immediately.

Cracking, crazing, discoloration and the like during the first week of operation are signs of severe 
chemical attack.  If this occurs, immediately remove the chemical from the pump.  Determine which 
parts are being attacked and replace them with parts that have been manufactured using more suitable 
materials.  The manufacturer does not assume responsibility for damage to the pump that has been 
caused by chemical attack.

7.2

How to Clean and Lubricate the A-100NV

The A-100NV will require occasional cleaning and lubricating.  The amount will depend on the severity 
of service.

]

When changing the pump tube assembly, clean the pump head chamber, TFD sensors, roller assembly 
and pump head cover.

]

The pump head cover bearing may require grease periodically.  Apply a small amount of grease 
(Aeroshell aviation grease #5 or equivalent) when necessary.

]

Silicone oil 

ONLY 

may be used on the roller assembly and tube assembly.

]

The injection/check valve assembly 

must be cleaned 

periodically, especially when injecting fluids that 

calcify such as sodium hypochlorite.  These lime deposits and other build ups can clog the fitting 
increasing the back pressure which can interfere with the check valve operation and possibly damage the 
fitting and/or the pump. See section 4.3.4.  Fig. 4.7.

]

Periodically clean the suction strainer.  Fig.4.5.

]

Periodically inspect the air vents located under the motor compartment and on the rear panel.  Clean if 
necessary.

7.3

500 Hour Service Warning Timer

The A-100NV is equipped with a service warning timer. After 500 hours of accumulated running time, 
the 

SERVICE 

icon will light. This is a reminder that the pump tube is nearing its minimum life expec-

tancy and should be replaced. 

Your actual tube life will depend on many factors such as the chemical 

used, back pressure, temperature, viscosity, and motor RPM.

Simultaneously press the 

RUN/STANDBY

 and 

DIGIT 

buttons to reset the service timer to zero. 

Note: 

Pressing the 

FIELD 

and 

DIGIT 

buttons will display the currently accumulated time value.

Page 16

Inlet Adapter

Outlet Adapter

Pump Head

FIG. 7.1

RUN

STANDBY

RUN

FIELD

DIGIT

MODE

PROGRAM

STAND-BY

PRIME

MINIMUM

MAXIMUM

INPUT MODES

1 - MANUAL

2 - 4-20mA

3 - 0-10VDC

4 - PULSE (Hz)

PROGRAM

RESET SERVICE

PRIME

DISPLAY

VARIABLE SPEED PUMP

% SPEED

1

MODE

VDC

SERVICE

1000

ALARM

mA

Hz

Содержание flexflo A-100NV

Страница 1: ...le Speed Peristaltic Injector Pump Operating Manual 5300 Business Drive Huntington Beach CA 92649 USA Phone 714 893 8529 FAX 714 894 9492 E mail sales blue white com or techsupport blue white com Webs...

Страница 2: ...o maintain the A 100N 16 7 1 Routine inspection and cleaning 16 7 2 How to clean and lubricate the A 100N 16 7 3 500 hour service warning timer 16 7 4 How to replace the pump tube 17 Replacement parts...

Страница 3: ...system 2 0 Specifications Maximum Working Pressure 100 psig 6 9 bar o o Maximum Fluid Temperature 130 F 54 C o o Ambient Temperature Range 14 to 110 F 10 to 43 C Duty Cycle Continuous Maximum Solids 5...

Страница 4: ...the pump tube Do not mount the pump directly over your chemical container Chemical fumes may damage the unit Mount the pump off to the side or at a lower level than the chemical container Mounting th...

Страница 5: ...umping unit R Suction Tube Chemical Container NPT Injector Page 5 RUN STANDBY RUN FIELD DIGIT MODE PROGRAM STAND BY PRIME MINIMUM MAXIMUM INPUT MODES 1 MANUAL 2 4 20mA 3 0 10VDC 4 PULSE Hz PROGRAM RES...

Страница 6: ...ections The A 100NV will accept any one of three different types of external input signals 4 20 mA 0 10 VDC or frequency Two types of frequency inputs AC sine waves magnetic coils type outputs and Dig...

Страница 7: ...WHITE frequency input JUNCTION BOX LIQUID TIGHT CONNECTOR ACCEPTABLE CABLE JACKET RANGE 118 255 INCH 3 0 6 5 MM EXTERNAL INPUT CABLE BROWN motor on contact output INPUT SIGNAL CONNECTOR 7 wire bundle...

Страница 8: ...e tube Secure the ceramic weight to the strainer Drop the strainer into the solution tank 4 3 3 Outlet Tubing Locate the outlet fitting of the Pump Tube see fig 4 6 Remove the tube nut Push the opaque...

Страница 9: ...he back pressure and interfering with the check valve operation See section 7 0 Install the Injection Check valve directly into the piping system Do not use a pipe stud with a tee for insertion of the...

Страница 10: ...pressed with the MODE button initiates the programming mode The ARROW next to the word PROGRAM will blink FIELD Button 4 In the programming mode selects the digit to be changed DIGIT Button 4 In the...

Страница 11: ...mode 2 Press the MODE button until MODE 2 is shown on the LCD display The SPEED or mA icon will light depending on the current display setting The large 3 DIGIT LCD will indicate the current motor spe...

Страница 12: ...LD button to select the digit to program The digit will blink when selected 4 Press the DIGIT button to change the selected digit 4 Repeat until all digits are programmed 4 Press the mode button Progr...

Страница 13: ...maximum VDC input signal value Press the FIELD button to select the digit to program The digit will blink when selected 4 Press the DIGIT button to change the selected digit 4 Repeat until all digits...

Страница 14: ...ut signal value Press the FIELD button to select the digit to program The digit will blink when selected 4 Press the DIGIT button to change the selected digit 4 Repeat until all digits are programmed...

Страница 15: ...in the bottom of the pump head just enough to cover the sensors and turn on the pump If the TFD system detects the chemical the pump will stop after a five second confirmation period and the ALARM ic...

Страница 16: ...may require grease periodically Apply a small amount of grease Aeroshell aviation grease 5 or equivalent when necessary Silicone oil ONLY may be used on the roller assembly and tube assembly The inje...

Страница 17: ...on side of the Pump Tube fitting into the lower retaining slot in the pump head Fig 7 2 6 4 2 2 Carefully guide the Pump Tube into the pump head Stretch the tube slightly and insert the outlet dischar...

Страница 18: ...0 15 16 11 3 14 13 31 30 17 26 27 3 2 1 2 9 52 54 5 4 7 8 56 55 10 12 28 29 25 24 23 22 21 33 32 34 F IE L D 6 53 D IG IT M O D E ST AN DB Y PR IM E R ES ET 99 SE C C YC LE TU BE LIF E WA RIN G TIM ER...

Страница 19: ...8 Star 1 25 90010 222 Wire Motor ground Digital Timers Green 1 26 A 008 1 Gearbox 14 Rpm 1 A 008 2 Gearbox 30 Rpm 1 A 008 3 Gearbox 45 Rpm 1 A 008 4 Gearbox 60 Rpm 1 A 008 5 Gearbox 125 Rpm 1 27 76001...

Страница 20: ...a brief description of the problem as well as the original invoice or sales receipt showing the date of purchase The receipt will be returned with the unit Prepay all shipping costs COD shipments will...

Отзывы: