G5 Weighing Instrument
1-5
Current output
4 – 20 mA, 0 – 20 mA, -12 – 20 mA, -20 – 20 mA,
< 500 ohm load. Current source, i.e. no external power supply
needed.
Update rate
300 Hz
Filter
Weight extra smoothing filter (on/off via set-up)
Fieldbus
ProfibusDP, DeviceNet and ControlNet.
Fieldbus
data
16 bytes from fieldbus to instrument.
80 bytes from instrument to fieldbus.
See chapter ‘Communication’ section ‘Fieldbus communication interface’ for
details on fieldbus data mapping.
Settings
All fieldbus settings are done with setup parameters in the instrument. No
settings are done on the module itself.
Mounting
The fieldbus adaptor is mounted in the fieldbus slot with LED’s and connector
accessible.
Remove the plastic cover from the fieldbus slot. Insert the adaptor very
carefully and make absolutely sure that the adaptor slides correctly into the
guides in the connector on the PCB. Tighten the two fastening screws at the
adaptor front and check that the two securing hooks locks into the PCB.
Module
type
Profibus-DP
DeviceNet
ControlNet
Connector Profibus 9-pin, female
D-sub (DB9F)
5 pin male connector.
BNC.
Baud rate
9.6 kbps – 12 Mbps or
Auto set by parameter
125, 250, 500 kbps or
Auto set by parameter.
Fixed 5 Mbit/s.
Address
1 – 125,
set by parameter
0 – 63,
set by parameter
1 – 99, set by parameter
Bus
Supply
Voltage
According to DeviceNet
(Node) specification:
nominal 24 VDC,
range 11 – 25 VDC.