G5 Weighing Instrument
3-9
Range/Alternatives
Explanation and
<default value>
result of alternatives.
Set-up Code
Range:
1 - 9999
<1937>
Defines the valid code for Set-up lock. If ‘Operator lock’ (see
above) is ‘On’ this code will still give access to all menus in the
Main menu.
Note: this parameter is only shown if ‘Set-up Lock‘ is set to ‘On’
Button-press Err. Msg.
Range:
Off
On
<Off>
For the application specific keys (ZERO, TARE, GROSS/NET
and PRINT) configured as ‘On’, this parameter is used when a
failed command shall generate a fail description message on the
user interface. The message will be shown for a few seconds.
Off: Do not show key press error messages. N.B. this will not
prevent WARNING messages from being show (see chapter
TBD).
On: Show key press error messages.
USB Memory Detect Time
Range:
1 - 6
Unit: s
<3>
Defines maximum detection time for USB memory at power on
and if there is no USB memory present this time will go to the
end. Some USB memory’s take longer time to react at power on
but to make power on time as short as possible this time can be
adjusted.
Gross After Print
Range:
No
Yes
<No>
After a Print command, display automatic change to Gross
weight.
No: No change after Print command.
Yes: Displaying Gross weight after Print command.
Warm Up Time
Range:
0 - 200
Unit: min
<0>
Defines the delay time in minutes from power up until the weight
presentation has full accuracy. Indicated in the display with the
text ‘Warming up!’.