34
Table 10. Ultrasound Acoustic Output Parameters (FDA Standard)
Values in this table are the maximum readings obtained from three test results
ACOUSTIC OUTPUT
MI
I
SPTA.3
(mW/cm
2
)
I
SPPA.3
(W/cm
2
)
Global Maximum
Value
0.268*
0.0977
8.06
Associated
Acoustic
Parameter
p
r.3
(MPa)
0.501
W
o
(mW)
0.339
0.339
f
c
(MHz)
3.54
3.54
3.54
z
sp
(cm)
2.90
2.90
Beam
dimensions
x
‑6
(cm)
0.306
y
‑6
(cm)
0.315
PD
(µsec)
0.658
0.658
PRF
(Hz)
400
400
EDS
Az. (cm)
4.75
Ele. (cm)
4.75
TIS/TIB/TIC range
0.0‑1.0*
*
Both MI and TI values are below 1.0.
CHARGING CRADLE SPECIFICATIONS
The charging cradle is tested to EN/IEC 60601‑1 requirements and is in compliance with UL and CSA
equivalent standards. The charging cradle is not intended for direct patient contact. It is designed to operate
within the specifications and environmental conditions identified in the following table.
Table 11.
Charging Cradle Specifications
GENERAL SPECIFICATIONS
Input Voltage
90–264 VAC RMS
Input Frequency
47–63 Hz
Input Current
0.5 Amp max
Input Connection
Direct plug‑in AC prongs for wall plug‑in units
Output
5V at 2.4 Amps
Insulation
Class II with double insulation
STORAGE SPECIFICATIONS
Storage
Indoor
Ambient Temperature Range
‑10°–50ºC (14°–122ºF)
Atmospheric Pressure Range
50–106 kPa
Relative Humidity
30–75% non‑condensing
Water Resistance
IPX0 (ordinary equipment without protection against ingress of water)