MAINTENANCE & TROUBLESHOOTING
REGULAR INSPECTIONS
Inspect the instrument for damage prior to each use. Inspect for cracks, abrasions, gouging, and evidence of
impact on the instrument. Cracks that allow the ingress of fluid may affect the safety and performance of
the instrument.
If you see any physical damage or cracks in the instrument, discontinue use immediately and contact
Verathon
®
Customer Care or your local representative.
IMPORTANT
CALIBRATING THE BLADDERSCAN INSTRUMENT
You must periodically calibrate your BladderScan instrument to make sure that it is providing accurate results.
The required frequency of calibration depends on your Total Reliability
SM
Plan. Calibrating the instrument on a
regular basis ensures accurate and proper alignment of the instrument’s internal coordinate system.
If you have the calibration kit and Scan Point
®
with QuickPrint, you can easily and quickly calibrate your own
instrument. You may also send your instrument to an authorized Verathon service center.
If you are not using Scan Point with QuickPrint, you must send your instrument in to an authorized Verathon
service center for calibration.
To contact Verathon Customer Care about calibration, please visit
verathon.com/support
.
PROCEdURE 1. CALIBRATE THE INSTRUMENT
Calibrating your BladderScan instrument ensures that it is providing accurate results, and maintains proper
alignment of the instrument’s internal coordinate system. If the instrument is not calibrated by the prescribed
date, the instrument is disabled and must be calibrated before it can be used again.
1. Place the calibration tank on a flat, non‑reflective surface, and then remove the lid.
2. Pour clean, room‑temperature water into the tank base, filling to the indicator mark. Ensure that there is
a minimal amount of bubbles in the water.
Note: The tank may need to sit for 24 hours until the water has degassed.