33
Operations & Maintenance Manual: Product Specifications
Table 9.
Ultrasound Acoustic Output Parameters (IEC Standard)
INDEX LABEL
MI
TIS
TIB
TIC
SCAN
NON‑SCAN
NON‑
SCAN
A
APRT
≤1 cm
²
A
APRT
>1 cm
²
Maximum index value
0.251 1.42E‑3
—
—
—
*
Associated
Acoustic
Parameter
p
r.a
(MPa)
0.470
P
(mW)
0.331
—
—
†
min of [P
α
(z
s
), I
ta,
α
(z
s
)] (mW)
—
z
s
(cm)
—
z
bp
(cm)
—
z
b
(cm)
—
z at max. I
pi,α
(cm)
2.60
d
eq
(z
b
)
(cm)
—
f
awf
(MHz)
3.50
3.50
—
—
—
†
Dim of A
aprt
X (cm)
0.957
—
—
—
†
Y (cm)
0.957
—
—
—
†
Other
Information
t
d
(μsec)
0.624
prr
(Hz)
400
p
r
at max. I
pi
(MPa)
0.644
d
eq
at max. I
pi
(cm)
—
I
pa.3
at max. MI
(W/cm
²
)
7.48
* Intended use does not include cephalic uses, so TIC is not computed.
† No data reported.
Notes:
1.
Information need not be provided for any formulation of TIS not yielding the maximum value of TIS for that mode.
2.
Information need not be provided regarding TIC for any transducer assembly not intended for transcranial or neonatal cephalic uses.
3.
Information on MI and TI need not be provided if the equipment meets both the exemption clauses given in 51.2 aa) and 51.2 dd).
4.
Scanning: Each frame of the overall 3D scan consists of scanning a sector and then rotating the sector plane about the center axis of the
scan through a full 360°. The maximum intensity values calculated only occur at the center of the rotation where the planes overlap.