Stim Switch User Manual
Rev. 2.00
®
Page v
LB-0357
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-‐4-‐11
30%/500ms
60%/100ms
100%/10ms
100%/5s
Test High & Low
Voltage @ 50 Hz only
30%/500ms
60%/100ms
100%/10ms
100%/5s
Mains power quality should be
that of a typical commercial or
hospital environment. If the
user of the Stim Switch requires
continued operation during
power mains interruptions, it is
recommended that the Stim
Switch be powered from an
uninterruptible power supply
or a battery.
Power Frequency
(50/60 Hz) magnetic
field
IEC 61000-‐4-‐8
3 A/m
Test both 50/60 Hz
3 A/m
It may be necessary to position
the Stim Switch further from
sources of power frequency
magnetic fields or to install
magnetic shielding. The power
frequency magnetic field
should be measured in the
intended installation location to
assure that it is sufficiently low.
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The Stim Switch is intended for use in the electromagnetic environment specified below. The customer
or the user of the Stim Switch should assure that it is used in such an environment
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment -‐
Guidance
Conducted RF
IEC 61000-‐4-‐6
Radiated RF
IEC 61000-‐4-‐3
3V
150kHz – 80MHz
80%am**
In ISM bands
a
3V/m
80-‐2500 MHz
80%am**
3V
3V/m
Portable and mobile RF
communications equipment
should be used no closer to any
part of the Stim Switch
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended Separation
Distance
d = 1,2
√
P 150kHz to 80MHz
d = 1,2
√
P 80MHz to 800MHz
d = 2,3
√
P 800MHz to 2.5GHz