 Stim Switch User Manual 

 

 

 
Rev. 2.00  

                                 ® 

Page 9 

LB-0357 

 

 

 

3

 

Software 

The StimComm software runs under Windows XP Professional, Windows Vista, or Windows 7. 
The StimComm software can be installed from the supplied CD by running the 

setup.exe

 

program.  Once the program has been installed, the StimComm software can be started by 
double clicking the StimComm desktop shortcut that was created by the installation program, or 
by launching the 

StimComm.exe

 program in the installation directory, 

C:\Program 

Files\Blackrock Microsystems\StimComm.

  The main features of the StimComm 

program are described below, following the screenshot of the program’s main window.  

NOTE:  Only open the StimComm Software after plugging in the headstage module(s) and 
turning on the power supply for the Stim Switch. Otherwise, the software will not recognize the 
headstage module. 

Electrode Selection Panel (1): 

Channels intended for stimulation can be selected by clicking the 

box to the left of the appropriate channel name so that a check appears in the box.  Electrodes 
can be independently selected for Banks A-D by selecting the Electrode tabs at the top of the 
window.  All 32 electrodes in the bank can be selected or deselected using the ‘Check All’ or 
‘Uncheck All’ buttons at the bottom of the panel.  Changes are not updated to the Stim Switch 
until the Set Electrodes button is pressed. 

Stim Control Panel (2):

  Stimulation and Headstage Module settings can be altered using the 

options that are available in the Stim Control panel: 

 

Disabled:

  In this mode, all stimulation channels for the selected bank are disconnected. 

Control On:

  This mode allows for continuous stimulation of the selected bank.  The                    

Control On feature can be used as a method for manually controlling switching between 

Figure 3-12: StimComm Main Window  

Содержание STIM

Страница 1: ...STIM SWITCH USER MANUAL REVISION 2 00 Blackrock Microsystems LLC 391 Chipeta Way Suite G Salt Lake City UT 84108 Tel 866 806 3692 www blackrockmicro com...

Страница 2: ...r Manual Rev 1 00 Page i LB 0357 Copyright 2011 Blackrock Micosystems LLC All rights reserved Copying or other reproduction of this document is prohibited without prior written consent of Blackrock Mi...

Страница 3: ...the physician surgeon considers at elevated risk of infection Electrocautery should not be performed while the Stim Switch is in use The Stim Switch should not be used if the device has been dropped D...

Страница 4: ...use Do not leave the Stim Switch connected to the patient electrodes when the system is not in use When connecting the Stim Switch to the Stim100 or patient electrodes use caution to minimize the like...

Страница 5: ...w and are not likely to cause any interference in nearby electronic equipment RF Emissions CISRP 11 Class B The Stim Switch is suitable for use in all establishments including domestic establishments...

Страница 6: ......

Страница 7: ...elding The power frequency magnetic field should be measured in the intended installation location to assure that it is sufficiently low Guidance and Manufacturer s Declaration Electromagnetic Immunit...

Страница 8: ...in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2 5 GHz are intended to decrease the likelihood that mobile portable communications equipment could cause in...

Страница 9: ...e after switching before action potentials are valid 5 ms using Fast Settle a modified Front End Amplifier 4 s using an unmodified Front End Amplifier Maximum Stimulation Input Current 30 mA on any on...

Страница 10: ...Care and Storage 11 5 Troubleshooting 12 6 Warranty 14 6 5 Return Merchandise Authorization 14 List of Figures Figure 1 1 Application Overview 1 Figure 2 2 Stim Switch Control Module Front Panel 2 Fig...

Страница 11: ...are recorded and stored by a data acquisition system The figure below shows an application overview of how a Stim Switch fits into a complete stimulation system 1 1 Packing List 1 Stim Switch Control...

Страница 12: ...vity on the port If an error occurs the LED will flash red FS Out 4 The Fast Settle Out BNC connector is used to drive the fast settling input of a NeuroPort Front End Amplifier The fast settling inpu...

Страница 13: ...nnectors 7 Each bank on the front panel of the Stim Switch contains a connector for a 50 conductor Headstage Module Cable These cables provide the interface between the Stim Switch and the Headstage H...

Страница 14: ...tion socket connector on the other Connections The DB37 connector connects to one of the 4 Stimulation Connectors on the front panel of the Stim Switch see Section 2 1 the 40 position connector connec...

Страница 15: ...r switch is only closed when the Headstage Module is configured in Stimulation mode and the channel has been selected for Stimulation using the StimComm software The table below defines the state of e...

Страница 16: ...in place using two thumbscrews 7 Plug the other end of the Headstage Cables into the available Headstage Modules that are encased within the Headstage Housing The Headstage Cables from each Bank A D...

Страница 17: ...n to the Data Acquisition System CAUTION The Stim Switch is ONLY approved for use with Blackrock Microsystems NeuroPort System 1 Verify that the Stim100 and the Stim Switch are powered off 2 Using a s...

Страница 18: ...o the electrode numbers and pressing the Set Electrodes button Electrode numbers can be selected independently for each bank of stimulation channels by selecting the tabs at the top of the window 4 Co...

Страница 19: ...software will not recognize the headstage module Electrode Selection Panel 1 Channels intended for stimulation can be selected by clicking the box to the left of the appropriate channel name so that...

Страница 20: ...not connected to either the stimulation source or the recording amplifier Set Stim The channels specified for stimulation in the Electrode Selection area are set to stimulation mode All channels not...

Страница 21: ...under the Port COM LPT heading in the device manager By default the StimComm program is set to use COM4 If you need to change the port assignment you can do that in the settings window After changing...

Страница 22: ...d case sensitive Type Blanking Mode to turn on the Blanking box 7 Right click on the newly added registry 8 Select Modify and change the 0 to 1 Leave the Base set to Hexadecimal If you want to turn th...

Страница 23: ...sent If the light doesn t light up or is red try disconnecting and then reconnecting the USB A B cable making sure that both ends are fully securely plugged into the ports 4 Check which COM ports are...

Страница 24: ...cular purpose In no event shall Blackrock Microsystems Inc be liable for any incidental or consequential damages or for the infringement of any patent rights or third party rights due to the use of it...

Отзывы: