Blackrock Microsystems STIM Скачать руководство пользователя страница 4

Stim Switch User Manual 

 

 

 

 
Rev. 2.00  

                                 ® 

Page iii 

LB-0357 

 

 

 

 

Use only the Front End Amplifier and Neural Signal Processor (NeuroPort™ System) supplied by 
Blackrock Microsystems.  Substitution of systems not supplied by Blackrock may affect system 
performance and patient safety. 

 

The Stim Switch and Amplifier/NSP should be disconnected from the patient electrodes during 
defibrillation.  The Stim Switch and its subcomponents, along with the Amplifier and NSP are not 
defibrillator proof. 

 

Do not use the Stim Switch in the presence of flammable anesthetic agents. 

 

Only connect Stim Switch components to properly tested, grounded and dedicated AC outlets to 
reduce the risk of electrical shock. Do not use an adapter for ungrounded wall outlets.  

 

Do not connect the Stim Switch to an outlet controlled by a wall switch. 

 

Avoid strong static discharges from sources like televisions or computer monitors because it can 
damage the electrical parts of the system. 

 

Keep the Stim Switch away from liquids. Contact with water, shower spray, or wet surfaces can lead 
to the patient receiving an electrical shock. 

 

Each component of the system must be plugged into a separate outlet (i.e. Stim Switch, Stim100, 
subcomponents of the data acquisition system and all other accessory equipment).  Using a power-
strip may cause leakage currents to exceed acceptable limits. 

 

Do not unplug the power supply to the Stim Switch while the system is in use. 

 

Do not leave the Stim Switch connected to the patient electrodes when the system is not in use. 

 

When connecting the Stim Switch to the Stim100 or patient electrodes, use caution to minimize the 
likelihood of Electrostatic Discharge (ESD). Do not connect/disconnect any cables to/from the Stim 
Switch while the system is in use. 

 

Ensure that the Stim Switch is securely positioned on a flat surface during use. 

 

Use caution when placing power cords, cables, and other connectors to minimize the likelihood of 
tripping or accidentally pulling on cables.  Damaged cables may results in failed ground 
connections. 

 

Connecting equipment to and from the Stim Switch may result in a summation of leakage currents 
that can lead to the patient receiving an electric shock. 

 

Risk of Electrical Shock Hazard:  Accessory equipment connected to the SIP/SOP port must be 
certified according to the respective IEC standards, i.e., IEC 60950 for data processing equipment 
or IEC 60601-1 for electromedical equipment in order to reduce the risk of electrical shock.    All 
combinations of equipment must be in compliance with IEC 60601-1-1 systems requirements.  
Anyone connecting additional equipment to the SIP/SOP port configures a medical system, and 
therefore is responsible that the system complies with the requirements of the system standard IEC 
60601-1-1. 

 

IEC 60950 approved Information Technology Equipment must be placed outside the “patient 
environment.”  The patient environment is defined as an area 1.5m (4.92 feet) from the patient. 

 

 

The following options are recommended if there is a need to remediate system leakage current:

 

Содержание STIM

Страница 1: ...STIM SWITCH USER MANUAL REVISION 2 00 Blackrock Microsystems LLC 391 Chipeta Way Suite G Salt Lake City UT 84108 Tel 866 806 3692 www blackrockmicro com...

Страница 2: ...r Manual Rev 1 00 Page i LB 0357 Copyright 2011 Blackrock Micosystems LLC All rights reserved Copying or other reproduction of this document is prohibited without prior written consent of Blackrock Mi...

Страница 3: ...the physician surgeon considers at elevated risk of infection Electrocautery should not be performed while the Stim Switch is in use The Stim Switch should not be used if the device has been dropped D...

Страница 4: ...use Do not leave the Stim Switch connected to the patient electrodes when the system is not in use When connecting the Stim Switch to the Stim100 or patient electrodes use caution to minimize the like...

Страница 5: ...w and are not likely to cause any interference in nearby electronic equipment RF Emissions CISRP 11 Class B The Stim Switch is suitable for use in all establishments including domestic establishments...

Страница 6: ......

Страница 7: ...elding The power frequency magnetic field should be measured in the intended installation location to assure that it is sufficiently low Guidance and Manufacturer s Declaration Electromagnetic Immunit...

Страница 8: ...in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2 5 GHz are intended to decrease the likelihood that mobile portable communications equipment could cause in...

Страница 9: ...e after switching before action potentials are valid 5 ms using Fast Settle a modified Front End Amplifier 4 s using an unmodified Front End Amplifier Maximum Stimulation Input Current 30 mA on any on...

Страница 10: ...Care and Storage 11 5 Troubleshooting 12 6 Warranty 14 6 5 Return Merchandise Authorization 14 List of Figures Figure 1 1 Application Overview 1 Figure 2 2 Stim Switch Control Module Front Panel 2 Fig...

Страница 11: ...are recorded and stored by a data acquisition system The figure below shows an application overview of how a Stim Switch fits into a complete stimulation system 1 1 Packing List 1 Stim Switch Control...

Страница 12: ...vity on the port If an error occurs the LED will flash red FS Out 4 The Fast Settle Out BNC connector is used to drive the fast settling input of a NeuroPort Front End Amplifier The fast settling inpu...

Страница 13: ...nnectors 7 Each bank on the front panel of the Stim Switch contains a connector for a 50 conductor Headstage Module Cable These cables provide the interface between the Stim Switch and the Headstage H...

Страница 14: ...tion socket connector on the other Connections The DB37 connector connects to one of the 4 Stimulation Connectors on the front panel of the Stim Switch see Section 2 1 the 40 position connector connec...

Страница 15: ...r switch is only closed when the Headstage Module is configured in Stimulation mode and the channel has been selected for Stimulation using the StimComm software The table below defines the state of e...

Страница 16: ...in place using two thumbscrews 7 Plug the other end of the Headstage Cables into the available Headstage Modules that are encased within the Headstage Housing The Headstage Cables from each Bank A D...

Страница 17: ...n to the Data Acquisition System CAUTION The Stim Switch is ONLY approved for use with Blackrock Microsystems NeuroPort System 1 Verify that the Stim100 and the Stim Switch are powered off 2 Using a s...

Страница 18: ...o the electrode numbers and pressing the Set Electrodes button Electrode numbers can be selected independently for each bank of stimulation channels by selecting the tabs at the top of the window 4 Co...

Страница 19: ...software will not recognize the headstage module Electrode Selection Panel 1 Channels intended for stimulation can be selected by clicking the box to the left of the appropriate channel name so that...

Страница 20: ...not connected to either the stimulation source or the recording amplifier Set Stim The channels specified for stimulation in the Electrode Selection area are set to stimulation mode All channels not...

Страница 21: ...under the Port COM LPT heading in the device manager By default the StimComm program is set to use COM4 If you need to change the port assignment you can do that in the settings window After changing...

Страница 22: ...d case sensitive Type Blanking Mode to turn on the Blanking box 7 Right click on the newly added registry 8 Select Modify and change the 0 to 1 Leave the Base set to Hexadecimal If you want to turn th...

Страница 23: ...sent If the light doesn t light up or is red try disconnecting and then reconnecting the USB A B cable making sure that both ends are fully securely plugged into the ports 4 Check which COM ports are...

Страница 24: ...cular purpose In no event shall Blackrock Microsystems Inc be liable for any incidental or consequential damages or for the infringement of any patent rights or third party rights due to the use of it...

Отзывы: