Redundant protective earth connection shall be made to either the medical 
device or the other equipment. 

 

The other equipment shall be powered from a medical grade (IEC 60601-1 
compliant) separating transformer. 

Cautions 

 

Federal law restricts this device to sale, by or on the order of a physician. 

 

Electromagnetic Compatibility 

The following tables document compliance levels and guidance from the IEC 60601-1-2 2007 Standard, for 
the electromagnetic environment in which the Stim Switch should be used.  Medical Electrical Equipment 
requires special precautions regarding EMC and should be installed and put into service according to the 
guidelines provided in the tables below. 

 

Guidance   and   Manufacturer’s   Declaration   –   Electromagnetic   Emissions   

The   Stim   Switch   is   intended   for   use   in   the   electromagnetic   environment   specified   below.      The   customer   
or   the   user   of   the   Stim   Switch   should   assure   that   it   is   used   in   such   an   environment   

Emissions   Test   

Compliance   

Electromagnetic   Environment   -­‐   Guidance   

RF   Emissions      
CISPR   11   

Group   1   

The   Stim   Switch   uses   RF   energy   only   for   its   internal   
function.      Therefore,   its   RF   emissions   are   very   low   and   
are   not   likely   to   cause   any   interference   in   nearby   
electronic   equipment.   

RF   Emissions      
CISRP   11   

Class   B   

The   Stim   Switch   is   suitable   for   use   in   all   
establishments,   including   domestic   establishments   and   
those   directly   connected   to   the   public   low-­‐voltage   
power   supply   network   that   supplies   buildings   used   for   
domestic   purposes.   

Voltage   Fluctuations/   
Flicker   Emissions   
   
IEC   61000-­‐3-­‐3   

Complies   

 

Guidance   and   Manufacturer’s   Declaration   –   Electromagnetic   Immunity

   

The   Stim   Switch   is   intended   for   use   in   the   electromagnetic   environment   specified   below.      The   customer   
or   the   user   of   the   Stim   Switch   should   assure   that   it   is   used   in   such   an   environment   

Immunity   Test   

IEC   60601   Test   Level   

Compliance   Level   

Electromagnetic   Environment   -­‐   

Guidance   

Electrostatic   
Discharge   (ESD)   
   
IEC   61000-­‐4-­‐2   

±2,   ±4,   ±6kV   Contact   
±2,   ±4,   ±8kV   Air   

±2,   ±4,   ±6kV   Contact   
±2,   ±4,   ±8kV   Air   

Floors   should   be   wood,   
concrete   or   ceramic   tile.      If   
floors   are   covered   with   
synthetic   material,   the   relative   
humidity   should   be   at   least   
30%.   

Electrical   Fast   
Transient/Burst   
   
IEC   61000-­‐4-­‐4   

AC-­‐   ±2kV   

DC,   signal,   control   

lines   -­‐   ±1kV   

Test   both   50/60   Hz   

AC-­‐   ±2kV   

DC,   signal,   control   

lines   -­‐   ±1kV   

Mains   power   quality   should   be   
that   of   a   typical   commercial   or   
hospital   environment.   

Surge   
   
IEC   61000-­‐4-­‐5   

Ac/dm   -­‐   ±0.5,   ±1kV   

Ac/cm   -­‐   ±0.5,   ±1,   ±2kV   

Test   both   50/50   Hz   

   

Ac/dm   -­‐   ±0.5,   ±1kV   

Ac/cm   -­‐   ±0.5,   ±1,   

±2kV   

   

Mains   power   quality   should   be   
that   of   a   typical   commercial   or   
hospital   environment.   

Содержание STIM

Страница 1: ...STIM SWITCH USER MANUAL REVISION 2 00 Blackrock Microsystems LLC 391 Chipeta Way Suite G Salt Lake City UT 84108 Tel 866 806 3692 www blackrockmicro com...

Страница 2: ...r Manual Rev 1 00 Page i LB 0357 Copyright 2011 Blackrock Micosystems LLC All rights reserved Copying or other reproduction of this document is prohibited without prior written consent of Blackrock Mi...

Страница 3: ...the physician surgeon considers at elevated risk of infection Electrocautery should not be performed while the Stim Switch is in use The Stim Switch should not be used if the device has been dropped D...

Страница 4: ...use Do not leave the Stim Switch connected to the patient electrodes when the system is not in use When connecting the Stim Switch to the Stim100 or patient electrodes use caution to minimize the like...

Страница 5: ...w and are not likely to cause any interference in nearby electronic equipment RF Emissions CISRP 11 Class B The Stim Switch is suitable for use in all establishments including domestic establishments...

Страница 6: ......

Страница 7: ...elding The power frequency magnetic field should be measured in the intended installation location to assure that it is sufficiently low Guidance and Manufacturer s Declaration Electromagnetic Immunit...

Страница 8: ...in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2 5 GHz are intended to decrease the likelihood that mobile portable communications equipment could cause in...

Страница 9: ...e after switching before action potentials are valid 5 ms using Fast Settle a modified Front End Amplifier 4 s using an unmodified Front End Amplifier Maximum Stimulation Input Current 30 mA on any on...

Страница 10: ...Care and Storage 11 5 Troubleshooting 12 6 Warranty 14 6 5 Return Merchandise Authorization 14 List of Figures Figure 1 1 Application Overview 1 Figure 2 2 Stim Switch Control Module Front Panel 2 Fig...

Страница 11: ...are recorded and stored by a data acquisition system The figure below shows an application overview of how a Stim Switch fits into a complete stimulation system 1 1 Packing List 1 Stim Switch Control...

Страница 12: ...vity on the port If an error occurs the LED will flash red FS Out 4 The Fast Settle Out BNC connector is used to drive the fast settling input of a NeuroPort Front End Amplifier The fast settling inpu...

Страница 13: ...nnectors 7 Each bank on the front panel of the Stim Switch contains a connector for a 50 conductor Headstage Module Cable These cables provide the interface between the Stim Switch and the Headstage H...

Страница 14: ...tion socket connector on the other Connections The DB37 connector connects to one of the 4 Stimulation Connectors on the front panel of the Stim Switch see Section 2 1 the 40 position connector connec...

Страница 15: ...r switch is only closed when the Headstage Module is configured in Stimulation mode and the channel has been selected for Stimulation using the StimComm software The table below defines the state of e...

Страница 16: ...in place using two thumbscrews 7 Plug the other end of the Headstage Cables into the available Headstage Modules that are encased within the Headstage Housing The Headstage Cables from each Bank A D...

Страница 17: ...n to the Data Acquisition System CAUTION The Stim Switch is ONLY approved for use with Blackrock Microsystems NeuroPort System 1 Verify that the Stim100 and the Stim Switch are powered off 2 Using a s...

Страница 18: ...o the electrode numbers and pressing the Set Electrodes button Electrode numbers can be selected independently for each bank of stimulation channels by selecting the tabs at the top of the window 4 Co...

Страница 19: ...software will not recognize the headstage module Electrode Selection Panel 1 Channels intended for stimulation can be selected by clicking the box to the left of the appropriate channel name so that...

Страница 20: ...not connected to either the stimulation source or the recording amplifier Set Stim The channels specified for stimulation in the Electrode Selection area are set to stimulation mode All channels not...

Страница 21: ...under the Port COM LPT heading in the device manager By default the StimComm program is set to use COM4 If you need to change the port assignment you can do that in the settings window After changing...

Страница 22: ...d case sensitive Type Blanking Mode to turn on the Blanking box 7 Right click on the newly added registry 8 Select Modify and change the 0 to 1 Leave the Base set to Hexadecimal If you want to turn th...

Страница 23: ...sent If the light doesn t light up or is red try disconnecting and then reconnecting the USB A B cable making sure that both ends are fully securely plugged into the ports 4 Check which COM ports are...

Страница 24: ...cular purpose In no event shall Blackrock Microsystems Inc be liable for any incidental or consequential damages or for the infringement of any patent rights or third party rights due to the use of it...

Отзывы: