Operations
47
11. Press the
key to select B level setting
TRANb
and press the
key.
For 2A, input 2 and Press the
key to confirm.
12. Press the
key to select B width setting
Tb
and press the
key.
For 0.003s, input .003 and Press the
key to confirm.
13. Press the
key to select transient operation mode
Tmode
and press the
key.
14. Select
Continuous
and press the
key to confirm.
15. Press the
key to exit.
16. Press the
+
to enter the
System
menu.
17. Press the
key to select
Trigger Source
and press the
.
18. Select
Manual
and press the
key to confirm.
19. Press the
key to exit.
20. Press the
key to turn on the electronic load’s input.
21. Press the
key on channel 1 module to enable transient operation.
22. Press the
key on mainframe panel to trigger transient operation
Transient Operation Programming Example
In remote mode, the following commands can be used to setup the same parameters used in the tutorial above. (refer
to MDL4U Series Programming Guide for more information).
CURRent:TRANsient:MODE CONTinuous
CURRent:TRANsient:ALEVel 10
CURRent:TRANsient:AWIDth 0.2ms
CURRent:TRANsient:BLEVel 2
CURRent:TRANsient:BWIDth 0.3ms
TRANsient ON
TRIGger:IMMediate