background image

2

Safety Summary

The following safety precautions apply to both operating and maintenance personnel and must be followed during all

phases of operation, service, and repair of this instrument.

Before applying power to this instrument:

• Read and understand the safety and operational information in this manual.

• Apply all the listed safety precautions.

• Verify that the voltage selector at the line power cord input is set to the correct line voltage. Operating the instrument

at an incorrect line voltage will void the warranty.

• Make all connections to the instrument before applying power.

• Do not operate the instrument in ways not specified by this manual or by B&K Precision.

Failure to comply with these precautions or with warnings elsewhere in this manual violates the safety standards of design,

manufacture, and intended use of the instrument. B&K Precision assumes no liability for a customer’s failure to comply

with these requirements.

Category rating

The IEC 61010 standard defines safety category ratings that specify the amount of electrical energy available and the

voltage impulses that may occur on electrical conductors associated with these category ratings. The category rating is

a Roman numeral of I, II, III, or IV. This rating is also accompanied by a maximum voltage of the circuit to be tested,

which defines the voltage impulses expected and required insulation clearances. These categories are:

Category I (CAT I):

Measurement instruments whose measurement inputs are not intended to be connected to the

mains supply. The voltages in the environment are typically derived from a limited-energy trans-

former or a battery.

Category II (CAT II):

Measurement instruments whose measurement inputs are meant to be connected to the mains

supply at a standard wall outlet or similar sources. Example measurement environments are portable

tools and household appliances.

Category III (CAT III):

Measurement instruments whose measurement inputs are meant to be connected to the mains

installation of a building. Examples are measurements inside a building’s circuit breaker panel

or the wiring of permanently-installed motors.

Category IV (CAT IV):

Measurement instruments whose measurement inputs are meant to be connected to the primary

power entering a building or other outdoor wiring.

Do not use this instrument in an electrical environment with a higher category rating than what is specified in this manual

for this instrument.

You must ensure that each accessory you use with this instrument has a category rating equal to or higher than the

instrument’s category rating to maintain the instrument’s category rating. Failure to do so will lower the category rating

of the measuring system.

Содержание 1696B Series

Страница 1: ......

Страница 2: ...so accompanied by a maximum voltage of the circuit to be tested which defines the voltage impulses expected and required insulation clearances These categories are Category I CAT I Measurement instrum...

Страница 3: ...r flammable atmosphere Do not operate the instrument in the presence of flammable gases or vapors fumes or finely divided particulates The instrument is designed to be used in office type indoor envir...

Страница 4: ...the covers are removed To avoid injuries always disconnect the power cord from the instrument disconnect all other connections for example test leads computer interface cables etc discharge all circui...

Страница 5: ...ove the instrument when a probe is connected to a circuit being tested Compliance Statements Disposal of Old Electrical Electronic Equipment Applicable in the European Union and other European countri...

Страница 6: ...erious injury indicates a hazardous situation which if not avoided will result in minor or moderate injury Refer to the text near the symbol Electric Shock hazard Alternating current AC Chassis ground...

Страница 7: ...put at Power Up 12 4 9 LCD Brightness 12 4 10 Enable Disable SCPI 12 5 Using Programming Features 13 5 1 Preset Program 13 5 2 Setting Timed Program 13 5 3 Run Timed Programming 13 6 Maintenance 14 6...

Страница 8: ...ggered and the output power will be cut off and FAULT warning will appear When you get this warning switch off the unit and remove all loading If you switch the unit back on again it should resume nor...

Страница 9: ...Fuse 4A Slow Blow 250V 2 3 Interface and Controls Item Description 1 Rotary control knob 2 Up Down keys 3 Dual function control keys 4 Negative output terminal black 5 Chassis ground terminal green 6...

Страница 10: ...current 2 Rotate control knob or press and to set voltage current level Press the control knob to toggle the cursor position 3 2 Setting Voltage and Current Using Keypad 1 Press to switch between set...

Страница 11: ...the Keypad and Rotary Knob 1 Press 2 Press within 3 seconds to unlock keypad and rotary knob 4 5 Setting the RS 485 Address 1 Press 2 Press within 3 seconds to enter into RS 485 address set menu 3 Us...

Страница 12: ...nds to enable output on power up 4 8 Disable Output at Power Up 1 Press 2 Press the down arrow key within 3 seconds to disable output on power up 4 9 LCD Brightness 1 Press 2 Press on numeric keypad w...

Страница 13: ...from 1 to 9999 cycles or infinity run You can run the unit through the sequence of subprograms for the input cycles number unless interrupted by pressing the CLEAR key 1 Press then press on the numer...

Страница 14: ...d then back on to see if the prob lem persists No output power Check the output for on off symbol on the display Press then Cannot get a high voltage setting within the rated maximum Check the upper v...

Страница 15: ...the Docs Software tab https www bkprecision com products power supplies 1687B 1 36v 0 10a dc power supply html See Figure 7 1 Figure 7 1 Download Link 1 Click CP210X_USB_Driver zip to begin the downl...

Страница 16: ...Multiple Power Supplies to PC with RS 485 Use the RS 485 interface at the back of the power supply to daisy chain up to 31 power supplies see Figure 7 9 The USB to RS 485 adapter shown in Figure 7 10...

Страница 17: ...e starts If there is already a saved connection the software will automatically connect to the instrument 7 4 1 Setup a new instrument Select the Setting tab See Figure 7 12 Click on Edit to access th...

Страница 18: ...Windows has assigned COM port 12 Click Save and the black display will light up to indicate a successful connection See Figure 7 15 If using RS 485 enter the three digit ID displayed on the RS 485 set...

Страница 19: ...et the number of cycles from 1 to 9999 and click to start or click to save the program into the instruments internal memory for future recall 7 4 4 External Timed Program External Timed Program is com...

Страница 20: ...an be used to view present or stored output data see Figure 7 24 Data can be saved and exported as CSV spreadsheet file or sent to a printer 7 4 7 Data Log Sampling Time and Voltage Upper Limit UVL Se...

Страница 21: ...PC Software Control and Installation 21 Figure 7 13 Figure 7 14 Figure 7 15...

Страница 22: ...PC Software Control and Installation 22 Figure 7 16 Figure 7 17 SCPI Selection Figure 7 18...

Страница 23: ...PC Software Control and Installation 23 Figure 7 19 Figure 7 20 Figure 7 21...

Страница 24: ...PC Software Control and Installation 24 Figure 7 22 Figure 7 23 Figure 7 24 Figure 7 25...

Страница 25: ...solution Voltage 10 mV Current 1 mA Power 1 mW Meter Accuracy Voltage Meter 1 2 counts for V 5 V Current Meter 1 2 counts for I 0 5 A Ripple Noise Voltage 30 mVp p 6 mVrms Current 10 mArms General Eff...

Страница 26: ...probes connectors and accessories that you are using with the device Customers not on an open account must include payment in the form of a money order or credit card For the most current repair charg...

Страница 27: ...ia our website www bk precision com Exclusions This warranty does not apply in the event of misuse or abuse of the product or as a result of unauthorized alterations or repairs The warranty is void if...

Отзывы: