BT-740 Operation manual
98
P/N : 740-ENG-OPM-EUR-D03
Bistos Co., Ltd.
2018.05
•
The monitor should be kept clean. It is recommended to regularly clean the
enclosure surface and the display screen. Cleaning the enclosure with non-
etching cleaner such as soap and water.
CAUTION
•
To avoid damaging the monitor:
Do not use strong solvents such as acetone.
Most cleaners must be diluted before use. Diluting should be according to
the manufacturer's instructions.
Do not use abrasive materials (such as steel wool).
Do not allow any liquid entering into the enclosure, and never immerse any
part of the device into liquid.
Do not leave any cleaning solution on the surface of any part of the device.
NOTE
Wipe the monitor and sensor surface with medical alcohol, dry it naturally or with
clean, dry, lint-free cloth.
Bistos is not liable for effectiveness of using these chemicals for infectious disease
control. Please consult the infectious disease control officers or experts of the
hospital for advice.
15.3 Disinfection
In order to avoid damage to the product, we recommend that the product is disinfected only
when it is deemed necessary by the hospital maintenance procedures. We also recommend
that the instrument to be disinfected must first be cleaned.
CAUTION
•
To prevent damage to the monitor, do not disinfect the monitor with gas (EtO) or
formaldehyde.