BT-740 Operation manual
107
P/N : 740-ENG-OPM-EUR-D03
Bistos Co., Ltd.
2018.05
18.3.2 NIBP
NIBP
Standards compliant
IEC 80601-2-30:2009
Measurement method
Automatic oscillometric method
Operating mode
Manual, automatic, continuous
Useful life
100, 000 times
Measurement interval in
automatic mode
1/2/3/4/5/10/15/30/60/90/120/180/240/480min
Typical measurement
time
20~40s
Normal mode measuring
range (mmHg)
Adult
Pediatric
Systolic blood pressure
40-270
40-200
Mean blood pressure
20-230
20-175
Diastolic blood pressure
10-210
10-162
Measurement accuracy
Maximum average error: ±5mmHg
Maximum standard deviation: 8mmHg
Resolution
1mmHg
Initial inflation pressure
Default
Pressure setting range
Adult
160mmHg
140mmHg, 160mmHg, 180mmHg
Pediatric
140mmHg
140mmHg, 160mmHg,
Overpressure protection
point (software)
Adult: 300mmHg
Pediatric: 240mmHg
Overpressure protection
point (hardware)
Adult: 320~330mmHg
Pediatric: 265~275mmHg
Static Pressure accuracy
±3mmHg
NIBP Alarm
Adult
Pediatric
NIBP upper limit (mmHg)
1 mmHg step
SYS
31 ~ 280
31 ~ 230
MAP
11 ~ 240
11 ~ 175
DIA
11 ~ 220
11 ~ 165
NIBP lower limit (mmHg)
1 mmHg step
SYS
30 ~ 279
30 ~ 229
MAP
10 ~ 239
10 ~ 174
DIA
10 ~ 219
10 ~ 164
NIBP Electrical characteristics
Supply voltage
10V~14V DC
Maximum power
consumption
3.6w
Quiescent current
50mA
Maximum current during
measurement
180mA
Maximum current during
inflation
300mA