7.3 Monitoring steps
7.3.1 Preparation
Before placing the electrode, prepare the patient’s skin in the following steps.
Skin preparation: Since the skin is a poor conductor, it is very important to treat the
patient's skin for electrode placement appropriately to make good contact between the
electrode and the skin. Select the flat position with less muscles for the electrode
placement, and refer to the method below for treatment of the skin:
Remove the body hair at the position for electrode placement.
Gently rub the skin at the position for electrode placement to remove dead skin cells.
Wash the skin thoroughly with soap and water (do not use ether and pure alcohol, as
this will increase the skin's impedance).
Dry the skin completely before placing the electrode.
Install the spring clip or stud prior to the placement of the electrodes.
Place the electrode on the patient.
Connect the ECG cable and ECG interface.
WARNING
Check if the lead is adequately attached and do not have any damage before
monitoring. When the ECG cable is unplugged, the screen will display “ECG Lead
Off” prompt, and trigger an audible and visual alarm.
7.3.2 Selecting lead
Select the ECG parameter area or wave area → “ECG Setup” menu
.
Select “Other Setup >>”→ “ECG Other Setup” menu
.
Select “Lead Type”, and select the ECG lead as needed.
“3-Lead”: 3-lead; ECG wave options: I, II, III.
“5-Lead”: 5-lead; ECG wave options: I, II, III, AVR, AVL, AVF, V.