BT-410F and BT-410A Operation manual

 

P/N:410-ENG-OPM-EUR-R12 

Bistos Co.,Ltd. 

2021,05 

0.4 General precautions on environment 

Do not keep or operate the equipment under the environment listed below. 

 

Avoid placing in an area 
exposed to moisture. Do not 
touch the Head worn light 
with wet hand. 

 

Avoid exposure to direct 
sunlight 

 

Avoid placing in an area 
where high variation of 
temperature exists. 
Operating temperature 
ranges from 10 ~ 40

Operating humidity ranges 
from 5 ~ 85 %. 

 

Avoid in the vicinity of 
electric heater.   

 

Avoid placing in an area 
where there is an excessive 
humidity rise or ventilation 
problem. 

 

Avoid placing in an area 
where there is an excessive 
shock or vibration. 

 

Avoid placing in an area 
where chemicals are stored 
or where there is in danger 
of gas leakage. 

 

Avoid dust and especially 
metal material enters into 
the Head worn light. 

 

Do not disjoint or 
disassemble the Head worn 
light. Bistos Co., Ltd. does not 
have liability of it. 

 

Power off when the Head 
worn light is not fully ready 
to operate. Otherwise, the 
Head worn light could be 
damaged.   

 

1 Introduction

 

1.1 General   

This chapter provides a general description of the BT-410F and BT-410A Head-worn light including: 

 

Brief Device Description 

 

Product Features 

 

Model Configurations 

1.2 Brief device description   

This Head-worn light is used for medical examination by fixing it on the head and illuminating with 
condensed light to the injured part. This unit doesn't intend to contact directly to the patient's body and 
can be used for all medical field except ophthalmology. All material is made of the light plastic, and is 
consist of main-body and headband. The main body is composed of a light source (LED), main frame, and 
battery pack. Optional accessories are consisting of an astral lamp for shadow less effect, and loupe 
(magnifying glass). The battery is a lithium-ion rechargeable one, and operating time is about 4 hours. Full 
recharge takes about 4 hours. LED's optical power will be a little decreased even though you use the unit 
normally according to recommended operating time. The expectedly possible operating time is for 
approximately 50,000 hours. 

 

1.3 Intended use   

The Head-worn light utilizes to provide supplemental light to improve visibility by illuminating during minor 
medical procedures such as examination. 
The device designed to use by a medical professionals by wearing it onto the head. 

 

1.4 Contraindication   

The head-worn light is not intended to the any procedure for ophthalmology. Do not glare the light directly 
to the patient’s eyes. It may cause temporary vision impairment. 

 

Содержание BT-410A

Страница 1: ...BT 410F and BT 410A Head worn light Operation Manual Keep this manual for future reference P N 410 ENG OPM EUR R12...

Страница 2: ...e However no responsibility is assumed by Bistos Corporation For its use or any infringements of patents or other rights of third parties that may result from its use No license is granted by implicat...

Страница 3: ...7 2 1 Product outlook and description 7 2 2 Loupe optional assembly 8 2 3 Astral lamp optional assemble 9 3 Operation 9 3 1 Adjust the headband 9 3 2 Turn on the power 9 3 3 Focusing the light 9 3 4...

Страница 4: ...n product label packaging and this manual in order to stand for the information about Stand by Those symbols identify safety information Warning and caution necessarily well know when operating the de...

Страница 5: ...old environment to a warm humid location Do not attempt to service the BT 410F and BT 410A Only qualified service personnel should attempt any needed internal servicing 0 3 Battery warnings WARNING Do...

Страница 6: ...rief Device Description Product Features Model Configurations 1 2 Brief device description This Head worn light is used for medical examination by fixing it on the head and illuminating with condensed...

Страница 7: ...0A has Fixed Focus lens but BT 410F has Variable focus lens BT 410F and BT 410A system consists of the following function and elements Model BT 410F BT 410A Control of Light Intensity X O Control of L...

Страница 8: ...ration 2 1 Product outlook and description Fig 2 1 BT 410A outlook and description Headband height adjustable part Headband part Headband size adjustable part Protect band Power On Off button LED indi...

Страница 9: ...e shown below Fig 2 3 Loupe Assembly 2 3 Astral lamp optional BT 410A only assembly The astral lamp can be assembled at the bottom of light source Remove the lens cap from the light source and connect...

Страница 10: ...e high charging voltage may cause the device malfunction 3 3 Focusing the light BT 410F only You can adjust the focal length between 25 30 cm Adjust the focus for the best light quality before treatme...

Страница 11: ...r is unseparated When charging unseparated be sure the device turned off To separate or assemble the battery pack refer below Fig 3 1 Battery pack assembly method Fig 3 2 Battery pack separation metho...

Страница 12: ...s during surgery or other cases 4 Maintenance and cleaning 4 1 Checking the light intensity It is recommended that the intensity of the light be checked before each use and at least every six months W...

Страница 13: ...h device or you have further questions please contact Bistos customer center 6 Specifications The BT 410F and BT 410A head worn light is classified as follows in accordance with EN IEC 60601 1 Categor...

Страница 14: ...ons flicker emissions IEC61000 3 3 Complies WARNING The BT 410F and BT 410A should not be used adjacent to or stacked with other equipment If adjacent or stacked use is necessary the device should be...

Страница 15: ...equency magnetic fields should be at levels characteristic of a typical commercial or hospital environment Conducted RF IEC61000 4 6 3 Vrms 150 kHz to 80 MHz 6 Vrms in ISM bands between 0 15 MHz and 8...

Страница 16: ...d with the following symbol IEC 60417 Ref no 5140 Non ionizing electromagnetic radiation Note 1 At 80 MHz and 800 MHz the higher frequency range applies Note 2 These guidelines may not apply in all si...

Страница 17: ...t is manufactured and passed through strict quality control and inspection Compensation standard concerning repair replacement refund of the product complies with Framework Act on Consumers noticed by...

Отзывы: