background image

BT-410F and BT-410A Operation manual

 

P/N:410-ENG-OPM-EUR-R12 

Bistos Co.,Ltd. 

2021,05 

11 

changed to orange light. When the charging has been finished, the light changes to green. 

 

2) Fault state 

In below case, the charger may be defected. 

 

There is no light or a flickering light. 

 

A red light still doesn’t change to orange. 

 

If a green light does not come on after sufficient time has been passed or if the light color changes from 

green to red or orange. 

 

 

CAUTION

 

The battery pack may be discharged when the device remains turned on for a long time. Please 
turn off the device when do not use it. 

Note

 

It is not necessary to charge the lithium-ion battery pack every day. Charging every day may 
decrease the life time of battery pack. Thus, please charge the battery pack when the LED 
indicator turns to orange. 

 

3.8 Optional operation 
1) Loupe 

This function can be used to enlarge the examination area. When not using the loupe, loupe can be 
separated or turn the loupe 90 degrees to secure the view. 

         

 

Fig. 3.4 Loupe turned 90 degrees

 

2) Astral Lamp (BT-410A only) 

This function can be used to reduce shadows during surgery or other cases. 

 

4 Maintenance and cleaning

 

4.1 Checking the light intensity 

It is recommended that the intensity of the light be checked before each use, and at least every six months. 
When required, please contact to your regional service agent or Bistos Co., Ltd. 

 

CAUTION

 

Only qualified service personnel should perform service, repair or intensity adjustment. 

 

4.2 Cleaning 

   

WARNING

 

Before cleaning, remove the battery pack. 

Remove dust from the exterior of the device with a soft brush or soft cloth wetted with tepid water. 
Remove the dust and dirt on the cable, wipe the unit with a cloth soaked in lukewarm water (40˚C / 104

). 

Also kindly wipe it with a clinical alcohol about once a week. 

 

CAUTION

 

  Do not use caustic or abrasive cleaners. 

  Do not clean with alcohol, acetone, or other solvents. 

  Never immerse the device or its parts. 

 

4.3 Battery disposal and handling 

Be cautious when disposing of the battery pack. Adhere to all applicable laws regarding recycling 
Where you try to keep the battery, avoid storing battery above 60°C (140°F). If clothing or skin comes in 
contact with material leaked from inside the battery, immediately wash with plenty of clean water. 
When you need to replace battery pack, please contact to your local distributor or Bistos Co., Ltd for 
technical support. 

 

Содержание BT-410A

Страница 1: ...BT 410F and BT 410A Head worn light Operation Manual Keep this manual for future reference P N 410 ENG OPM EUR R12...

Страница 2: ...e However no responsibility is assumed by Bistos Corporation For its use or any infringements of patents or other rights of third parties that may result from its use No license is granted by implicat...

Страница 3: ...7 2 1 Product outlook and description 7 2 2 Loupe optional assembly 8 2 3 Astral lamp optional assemble 9 3 Operation 9 3 1 Adjust the headband 9 3 2 Turn on the power 9 3 3 Focusing the light 9 3 4...

Страница 4: ...n product label packaging and this manual in order to stand for the information about Stand by Those symbols identify safety information Warning and caution necessarily well know when operating the de...

Страница 5: ...old environment to a warm humid location Do not attempt to service the BT 410F and BT 410A Only qualified service personnel should attempt any needed internal servicing 0 3 Battery warnings WARNING Do...

Страница 6: ...rief Device Description Product Features Model Configurations 1 2 Brief device description This Head worn light is used for medical examination by fixing it on the head and illuminating with condensed...

Страница 7: ...0A has Fixed Focus lens but BT 410F has Variable focus lens BT 410F and BT 410A system consists of the following function and elements Model BT 410F BT 410A Control of Light Intensity X O Control of L...

Страница 8: ...ration 2 1 Product outlook and description Fig 2 1 BT 410A outlook and description Headband height adjustable part Headband part Headband size adjustable part Protect band Power On Off button LED indi...

Страница 9: ...e shown below Fig 2 3 Loupe Assembly 2 3 Astral lamp optional BT 410A only assembly The astral lamp can be assembled at the bottom of light source Remove the lens cap from the light source and connect...

Страница 10: ...e high charging voltage may cause the device malfunction 3 3 Focusing the light BT 410F only You can adjust the focal length between 25 30 cm Adjust the focus for the best light quality before treatme...

Страница 11: ...r is unseparated When charging unseparated be sure the device turned off To separate or assemble the battery pack refer below Fig 3 1 Battery pack assembly method Fig 3 2 Battery pack separation metho...

Страница 12: ...s during surgery or other cases 4 Maintenance and cleaning 4 1 Checking the light intensity It is recommended that the intensity of the light be checked before each use and at least every six months W...

Страница 13: ...h device or you have further questions please contact Bistos customer center 6 Specifications The BT 410F and BT 410A head worn light is classified as follows in accordance with EN IEC 60601 1 Categor...

Страница 14: ...ons flicker emissions IEC61000 3 3 Complies WARNING The BT 410F and BT 410A should not be used adjacent to or stacked with other equipment If adjacent or stacked use is necessary the device should be...

Страница 15: ...equency magnetic fields should be at levels characteristic of a typical commercial or hospital environment Conducted RF IEC61000 4 6 3 Vrms 150 kHz to 80 MHz 6 Vrms in ISM bands between 0 15 MHz and 8...

Страница 16: ...d with the following symbol IEC 60417 Ref no 5140 Non ionizing electromagnetic radiation Note 1 At 80 MHz and 800 MHz the higher frequency range applies Note 2 These guidelines may not apply in all si...

Страница 17: ...t is manufactured and passed through strict quality control and inspection Compensation standard concerning repair replacement refund of the product complies with Framework Act on Consumers noticed by...

Отзывы: