BT-300
Operation Manual 4
P/N : 300-ENG-OPM-EUR-R09
Bistos Co.,Ltd.
2018,03
Section
1
Safety
1.1 Instructions for the Safe Operation and Use of the
BT-300 Monitor
Examine the monitor and any accessories periodically to ensure that
the cables, line cords, transducers, and instruments do not have visible
evidence of damage that may affect patient safety or monitoring
performance. The recommended inspection interval is once per week
or less. Do not use the monitor if there is any visible sign of damage.
Only the AC line cord supplied with the BT-300, or its equivalent, is
approved for use with the Unit.
Do not attempt to service the BT-300 monitor. Only qualified service
personnel by Bistos Co., Ltd. should attempt any needed internal
servicing.
The BT-300 is not specified or intended for operation during the use of
defibrillators or during defibrillator discharge.
The BT-300 is not specified or intended for operation in the presence
of electrosurgical equipment.
The BT-300 is not specified or intended for operation in conjunction
with any other type of monitoring equipment except the specific
devices that have been identified for use in this Operator’s Manual.
Perform periodic safety testing to insure proper patient safety. This
should include leakage current measurement and insulation testing.
The recommended testing interval is once per year.
Do not operate the BT-300 monitor if it fails to pass the power on self-
test procedure.
WARNING
: Be informed that it may cause serious injury or death to the
patient, property damage, and material losses against the “Warning” sign.
CAUTION
: Be informed that it may cause no harm in life but lead to
injury against the “Caution” sign.