BT-300
Operation Manual 17
P/N : 300-ENG-OPM-EUR-R09
Bistos Co.,Ltd.
2018,03
The limit values are configurable. These limit values have no significant
meaning in clinical uses. To prevent overwrapping of limit value, there is an
apartness of upper or lower limit by 5 bpm. The purpose of setting for the limit
values is to give accommodation to user. But the delay from onset to alert is
fixed to 20 seconds. If alert situation is continued over 20 seconds, an alarm
event results in an audible tone.
Pressing the alarm button on the monitor’s keypad can silence the alarm tone.
Alarms are enabled or disabled by pressing the alarm on/off button. If alarms are
disabled then all alarms are off. If alarms are enabled then all alarms are on. The
following section describes the procedure used to set alarm parameters for
ultrasound heart rates.
Activity
Desired Result
[
]Button Press
To enter the setup menu over 2 seconds.
[
] FHR1Button Press
To select “UPPER LIMIT,” “LOWER LIMIT.”
[
] FHR2Button Press
To change the desired value.
The list below shows the values that are available
for each parameter:
Heart Rate Upper Limit {(Heart Rate Lower Limit
+10) ~ 235 BPM, 1 BPM increments}
Heart Rate Lower Limit {35 ~ (Heart Rate Upper
Limit-5) BPM, 1 BPM increments}
[
]Button Press
To save and exit setup menu.
4.1.3
Setting [Print Speed / Fetal Movement Print / Auto Print / Dop2
Offset / Record Enlarge]
This section describes the procedure used to set the paper speed and auto stop
time.
When ultrasound trace separation is enabled, the trend data for ultrasound
channel 2 is shifted up by 20 BPM in printing. This feature is provided to clearly
see separate heart rate trends when both heart rates are similar. The heart rate
value shown in the numeric frame is not affected. When fetal movement is
enabled, fetal movement graph is printed. FM1 graph is printed in upper
area(50~100 in UC graph Frame), and FM2 graph is in lower area(0~50 in UC
graph Frame). When follow the steps below to change the US graph separation
setting.
Activity
Desired Result
[
]Button Press
To enter the setup menu over 2 seconds.
[
] FHR1Button Press
To select “Print Speed”, “Fetal Movement Print”,
“Auto Print”, “DOP2 OFFSET”, “Record
Enlarge”
[
] FHR2Button Press
The list below shows the values that are available
for each parameter:
Print Speed {10, 20, 30 mm/minute}
Fetal Movement Print {ON or OFF}
Auto Print{0, 10, 20, 30, 40, 50, 60 minute}
Dop2 Offset {0 BPM,” or “20 BPM.}
Record Enlarge {ON or OFF}
[
]Button Press
To save and exit setup menu.