BT-300
Operation Manual 29
P/N : 300-ENG-OPM-EUR-R09
Bistos Co.,Ltd.
2018,03
Section
9
Cleaning and Disinfection
This chapter contains instructions for the care and cleaning of the BT-300 unit and its
accessories.
The BT-300 requires proper care and preventive maintenance. This ensures consistent
operation and maintains the high level of performance necessary in monitoring
procedures.
9.1
Monitor
Keep the external surface clean and free of dust, dirt, and residual liquids.
Clean with a damp cloth using mild soap and water or hospital approved,
nonabrasive disinfectants.
9.2
Transducers
Cleaning and Disinfecting the Tocotonometer and Ultrasound
Transducer
To avoid damage to the transducers, clean and disinfect only according to the
following instructions. Care MUST be taken to preserve both the
Tocotonometer label and the Tocotonometer cable label. DO NOT remove,
conceal or deface Tocotonometer labels.
1. Wipe the device with a sterile wipe soaked in enzymatic detergent safe for
use with metal instruments. Wipe the exterior of the device three times.
Prepare the detergent according to the manufacturer’s transducer
recommendations.
2. Scrub the transducer with enzymatic detergent using soft bristled brush for
five (5) minutes.
.
3. Wipe the transducer three (3) times with sterile water to remove soap
residue.
4. Wipe the transducer with a sterile wipe soaked in Cidex™. Wipe all
exterior surfaces of the transducer three (3) times.
5. Wipe the transducer three (3) times with sterile water to remove Cidex
WARNING
: Unplug the monitor from the AC power source and detach all
accessories before cleaning. Do not immerse the unit in water or allow liquids to
enter the case.
CAUTION
: Take extra care when cleaning the display surfaces, which are
sensitive to rough handling. Rub the lens that covers them with a soft, dry cloth.
CAUTION
: Do not autoclave. Does not gas or sterilize.
CAUTION
: Do not immerse in liquid. When using solutions, use sterile
wipes to avoid pouring fluids directly on the transducer.