20
10. SEALING ACCESS TO THE BALANCE SETTINGS
For legal for trade applications, the balance must be sealed to prevent access to the metrological parameters.
To seal the balance, press the recessed Lock switch at the bottom of the balance, momentarily during power up (the
message “LEGAL” appears an instant), then cover the Lock switch access hole. A cover seal and paper seal may be
installed as shown.
To regain access to the locked balance settings, break the seal and press the recessed Lock switch
momentarily during power up (the message “HIRES” appears an instant).
11. ERROR MESSAGES
ERROR
MESSAGES
DESCRIPTION
RESOLUTION
- - oL - -
Over range
Remove weight from the scale.
Err 4
Zero Setting Error
Upon start-up or when the
ZERO
key is pressed, the weight on the
scale exceeds the percentage programmed on the max. capacity.
Remove the excess weight and retry.
Err 5
Keyboard error
The keyboard could be damaged.
Err 6
Input signal greater
than 3mV/V
Indicates that the A/D converter value is greater than the maximum
value:
- Remove the weight from the scale if it’s exceeding.
- The load cell or the electronics could be damaged.
RECYCLING
INSTRUCTION
The crossed-out wheeled bin on the product means that at the product end of life, it must be taken
to separate collection or to the reseller when a new equivalent type of equipment is purchased. The
adequate differentiated refuse collection in having the product recycled, helps to avoid possible
negative effects on the environment and health and supports the recycling of the materials of which
the equipment is made. The unlawful disposal of the product by the user will entail fines foreseen
by the current regulations.