14
“F0” - CALIBRATION (*)
Premise
: The unit of measure of the calibration is fixed at kilogram (“kg”).
1)
The displays show “CAL F0”: press the
TARE
key to get into the step.
2)
The displays show the acceleration value set in the F16 step: enter the gravitational acceleration value of the
calibration zone
if different from the
use zone
and press the
TARE
key to confirm;
one must modify all the 6
digits of the gravitational acceleration.
The value will be set also in the F16 step.
3)
The display shows “UnLoAd”: unload the scale and press
TARE
.
4)
The displays show “LoAd X KiLoS”, in which X is the sample weight to be put on the scale; press several time the
M+
key if one wants to change the calibration weight.
5)
Put the weight on the scale, please wait for the weight stability and press
TARE
.
6)
If the calibration is correct, the displays show for an instant “SPAN PASS”, and then “CAL F0”.
If the calibration is uncorrect, the displays show for an instant “SPAN FAIL LOW” (if the weight is too low) or
“SPAN FAIL HI” (if the weight is too high) and then “CAL F0”.
IN CASE THE ZONE OF USE IS DIFFERENT FROM THE CALIBRATION ZONE ONE SHOULD:
1)
Carry out the calibration as previously described.
2)
Enter the
F16
step and set the gravity acceleration value for the ZONE OF USER.
3)
Press the
TARE
key to confirm
4)
Save and exit the SET-UP environment (press many times the
ZERO
key).
5)
The weight error, caused by the different gravitational attraction value between the calibration zone and the zone of
use, is corrected automatically.
“F1” - NUMBER OF DIVISIONS (*)
By pressing
TARE
one selects the scale divisions number.
Press
M+
key to select one value of the following:
Value
Divisions
1
3000
3000
1
6000
6000
dUAL rAnGE
3000 + 3000
“F2” - MAIN PRICE UNIT
By pressing
TARE
one selects the main price unit.
Press
M+
to select one of the possible unit:
kiLoS
kilogram
100
100 grams
“F3” - SCALE CAPACITY (*)
By pressing
TARE
one selects the total capacity of the scale.
Press
M+
key to select one of these capacities: 3, 6, 15, 30 kg.
“F4” - NUMBER OF SCALE DECIMALS (*)
By pressing
TARE
one selects the number of scale decimals.
Press
M+
to select one of the suggested values: 0.0 (a decimal), 0.00 (two decimals), 0.000 (three decimals), 0.0000
(four decimals), 0 (no decimal); confirm with
ZERO
.
“F5” - NUMBER OF AMOUNT DECIMALS
By pressing
TARE
one selects the number of amount decimals.
Press
M+