Appendix
42
Electromagnetic emitted interference according to IEC 60601-1-2
The CardioMessenger is suitable for operation in the indicated electromagnetic environment.
The customer and/or operator of the CardioMessenger should make sure that it is used in an electro-
magnetic environment as described below.
Checking
immunity to noise
Test level according
to IEC 60601-1-2
Conformance
level
Guidelines for the electromagnetic environment
Discharge of static
electricity (ESD)
According to
IEC 61000-4-2
± 6 kV contact discharge
±8 kV air discharge
Same as test level
Floors should be made of wood or cement, or have ceramic
tiles. When the floor consists of a synthetic material, the
relative humidity must be at least 30%.
Rapid transient
electrical disturbances/
bursts
According to
IEC 61000-4-4
± 2 kV for
power supply lines
± 1 kV for input/output
lines
Same as the test
level
The quality of the supply voltage should correspond to that in a
typical business and/or hospital.
Surge
According to
IEC 61000-4-5
±1kV normal mode
voltage
± 2 kV common mode
voltage
Same as the test
level
The quality of the supply voltage should correspond to that in a
typical business and/or hospital.
CM-S_TLine.book Page 42 Wednesday, December 21, 2011 4:58 PM
Содержание CardioMessenger-S TLine
Страница 4: ...2 CM S_TLine book Page 2 Wednesday December 21 2011 4 58 PM ...
Страница 34: ...Precautionary measures 32 CM S_TLine book Page 32 Wednesday December 21 2011 4 58 PM ...
Страница 50: ...CM S_TLine book Page 48 Wednesday December 21 2011 4 58 PM ...
Страница 51: ...371017 B_GA_CardioMessenger S_TLine_en US_Cover indd 3 20 12 2011 19 08 26 ...