CardioMessenger-S TLine
Appendix
43
Voltage dips, brief
interruptions and
fluctuations in the
supply voltage
According to
IEC 61000-4-11
‹
5% U
T
for
1/2 cycle
(
›
95% drop)
40% U
T
for 5 periods
(60% drop)
70% U
T
for 25 periods
(30% drop)
‹
5% U
T
for 5 s
(
›
95% drop)
Same as the test
level
The quality of the supply voltage should correspond to that in a
typical business and/or hospital.
Magnetic field at the
supply frequencies
(50/60 Hz)
According to
IEC 61000-4-8
3A/m
Same as the test
level
The magnetic field strength should correspond to the typical
value in business and hospital environments.
Comment: U
T
is the mains alternating voltage before applying the test levels.
Checking