Test Procedure
1. If you have not already done so, prepare the solutions for the tests you plan to
perform. See instructions starting on page
Refer to the pipette maps on the next few pages for the remaining steps.
2. Perform the Corners/Sensitivity/Linearity tests using the Top optics of the filter-
based fluorescence system:
l
Pipette the solutions into a new 96-well solid black or glass-bottom plate.
l
Create an experiment based on
Synergy H1_FI_T_SF.prt
and read the plate.
3. If your reader is equipped with Fluorescence Polarization capability:
l
Pipette the solutions for the “FP” test into the same plate as used in step 2.
l
Create an experiment based on
Synergy H1_FP.prt
and read the plate.
4. If your reader is equipped with the monochromator-based fluorescence system,
perform the Sensitivity/Linearity tests for that system:
l
Create an experiment based on
Synergy H1_M_FI_B_SF.prt
(bottom optics)
and read the plate.
l
Create an experiment based on
Synergy H1_M_FI_T_SF.prt
(top optics) and
read the plate.
5. If your reader is equipped with Time-Resolved Fluorescence capability:
l
Pipette the solutions for the “TRF” test into a new 96-well solid white plate.
l
Create an experiment based on
Synergy H1_TRF.prt
and read the plate.
6. Save the experiments. Refer to the instructions starting on page
to perform
calculations and determine pass/fail.
l
Troubleshooting tips are provided on page
.
l
Test descriptions are provided on page
.
126 | Chapter 8: Instrument Qualification Procedures
BioTek Instruments, Inc.
Содержание Synergy H1
Страница 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Страница 2: ......
Страница 26: ...BioTek Instruments Inc xxiv Preface ...
Страница 52: ...Figure 2 9 Packing the dispense module accessories 26 Chapter 2 Installation BioTek Instruments Inc ...
Страница 53: ...Figure 2 10 Packing the dispense module Synergy H1 Operator s Manual Repackaging and Shipping Instructions 27 ...
Страница 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Страница 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Страница 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...
Страница 180: ...Figure C 1 Top view 152 Appendix C Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...