l
Emission filter 528/20 nm installed
l
510 nm dichroic mirror and polarizers installed
Time-Resolved Fluorescence (TRF) Test:
BioTek offers a pre-configured qualification TRF filter cube for purchase;
contact BioTek Customer Care and ask about part number 8040555.
l
The recommended test solution (FluoSpheres carboxylate-modified microspheres,
0.2 μm europium luminescent, 2 μL) is available from Invitrogen Corporation
(#F20881) or BioTek (#7160011; see page
)
l
A new, clean 96-well solid white microplate, such as Corning Costar #3912
l
15-mL conical-bottom, polypropylene sample tube
l
Excitation filter 360/40 nm installed
l
Emission filter 620/40 nm installed
l
400 nm dichroic mirror installed
Test Solutions
Determine which tests to run for your reader model. Prepare the necessary test solutions
using the materials described on the previous pages.
Corners/Sensitivity/Linearity (FI) Tests:
If using BioTek’s sodium fluorescein powder (BTI #98155), be sure to
hold the vial upright and open it carefully; the material may be
concentrated at the top. If a centrifuge is available, spin down the tube
before opening.
When diluting the sodium fluorescein powder in buffer, it takes time for
the powder to completely dissolve. Allow the solution to dissolve for five
minutes, with intermittent vortexing, before preparing the titration dyes.
Wrap the vial containing the stock solution in foil to prevent exposure to
light. Discard unused solution after seven days. Discard any open,
unused buffer solution after seven days.
1. The Sodium Borate solution does not require further preparation; proceed to step 2.
If you are using PBS, prepare the solution:
l
(Optional, but recommended) Using a 0.45-micron filter, filter 200 mL of
deionized or distilled water.
l
Follow the manufacturer’s instructions on the PBS packaging to create 200 mL,
dissolving the necessary amount of PBS into the filtered water.
124 | Chapter 8: Instrument Qualification Procedures
BioTek Instruments, Inc.
Содержание Synergy H1
Страница 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Страница 2: ......
Страница 26: ...BioTek Instruments Inc xxiv Preface ...
Страница 52: ...Figure 2 9 Packing the dispense module accessories 26 Chapter 2 Installation BioTek Instruments Inc ...
Страница 53: ...Figure 2 10 Packing the dispense module Synergy H1 Operator s Manual Repackaging and Shipping Instructions 27 ...
Страница 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Страница 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Страница 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...
Страница 180: ...Figure C 1 Top view 152 Appendix C Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...