background image

11 

6.  Specifications 

The unit is designed for operation in cold rooms and closed laboratory rooms at am-

bient temperature from +4°C to +40°C in a non-condensing atmosphere and maximum rel-
ative humidity 80% for temperatures up to 31°C decreasing linearly to 50% relative humidity 
at 40°C. 

Biosan is committed to a continuous programme of  improvement and reserves the 

right to alter design and specifications of the equipment without additional notice. 

 

LMC-3000 

LMC-4200R 

Speed setting 

range 

In RPM 

100 

– 3000 RPM 

100 

– 4200 RPM 

In RCF

1

 

10 

– 1610 g 

10 

– 3160 g 

Speed increment 

100 RPM or 10 g 

Timer setting 

– 90 min 

Timer increment 

1 min 

Acceleration, 

RPM/s 

Slow mode 

50 

50 

Normal mode 

75 

75 

Fast mode

2

 

135 

150 

Deceleration, 

RPM/s 

Brakes off

2

 

30 

30 

Slow mode 

Normal mode 

45 

50 

Fast mode

2

 

135 

150 

Temperature setting range 

-

10 °C … +25 °C 

Stable temperature 

maintenance range 

25 °C below ambient … 

+25 °C 

Temperature setting resolution 

1 °C 

Rotor imbalance automatic diag-

nostics 

Emergency stop, dis

play indication “IMBALANCE” 

Rotation direction 

Counterclockwise 

Display 

2x16 characters, LCD 

2x24 characters, LCD 

Noise level 

≤ 60 dBA 

≤ 65 dBA 

Chamber diameter 

340 mm 

360 mm 

Dimensions 

495x420x235 mm 

635x580x335 mm 

Power consumption 

110 W (

V.6AD

); 

120 W (

V.6AE

990 W 

Operational voltage 

230 V, 50/60 Hz (

V.6AD

); 

120 V, 50/60 Hz (

V.6AE

230 V, 50/60 Hz 

Weight

3

 

11.8 kg 

56 kg 

 

 

1

 Relative Centrifugal Force 

2

 Depending on the load 

3

 

Accurate within ±10% 

Содержание LMC-3000

Страница 1: ......

Страница 2: ...6 Specifications 11 7 Ordering information 12 8 Care and maintenance 13 9 Warranty 14 10 EU Declaration of conformity 15 1 About this edition of user instructions The current edition of the user inst...

Страница 3: ...hin a 300 mm area around the device during the centrifuge operation Use only original parts and accessories provided by manufacturer for this product Before using any cleaning or decontamination metho...

Страница 4: ...4 3 Do not leave the operating unit unattended Do not operate the unit in environments with aggressive or explosive chemical mix tures Please contact manufacturer for possible operation of the unit i...

Страница 5: ...pounds etc and material itself blood components CSF and other thermolabile materials Temperature control absence at this stage causes unpredictable results FEATURES User friendly centrifugation parame...

Страница 6: ...ench for rotor fixation 1 pcs Emergency opening tool for LMC 4200R 1 pce Operating manual declaration of conformity 1 copy 4 2 2 Optional accessories R 6 rotor on request R 6P rotor on request R 12 10...

Страница 7: ...wear or corrosion and replace if necessary Hold the rotor with one hand and using the supplied wrench for rotor replacement 13 mm turn the fixation nut fig 1 1 counter clockwise to release the rotor...

Страница 8: ...tor on the display see 5 4 3 Figure 4 LMC 3000 control panel Figure 5 LMC 4200R control panel 5 1 Check the power cable for any signs of damage Connect the power cable to a properly grounded mains soc...

Страница 9: ...eration speed values can be found in the Specifications section This information can also be found on the lid of the unit 5 4 3 Installed rotor fig 4 3 or 5 3 Select the rotor that is currently instal...

Страница 10: ...leration mode press and hold RUN STOP key for more than 2 seconds 5 9 Press the Open key and open the lid by lifting it upwards with your hand It is possible to unlock and open the lid only when the r...

Страница 11: ...Timer increment 1 min Acceleration RPM s Slow mode 50 50 Normal mode 75 75 Fast mode2 135 150 Deceleration RPM s Brakes off2 30 30 Slow mode 8 8 Normal mode 45 50 Fast mode2 135 150 Temperature settin...

Страница 12: ...0208 BK R 24 10 For 24 round bottom centrifuge tubes of 10 15 ml xH 16x90 mm N A 4000 RPM 2860g BS 010212 JK R 2 For 2 standard 96 well microplates LxWxHmax 128x85 6x45 mm 2000 RPM 560g 2000 RPM 560g...

Страница 13: ...n of the unit 8 3 2 Disinfect the unit after operation by cleaning the parts inside the centrifuge chamber The rotor and other accessories are autoclavable 120 20 min 8 3 3 For LMC 4200R Wipe away any...

Страница 14: ...200R the Premium class model one year of extended warranty is availa ble free of charge after registration during 6 months from the date of sale Online registration form can be found in section Warran...

Страница 15: ...EU Applicable Standards LVS EN 61326 1 2013 Electrical equipment for measurement control and laboratory use EMC requirements General requirements LVS EN 61010 1 2011 Safety requirements for electrical...

Страница 16: ...Biosan SIA Ratsupites 7 build 2 Riga LV 1067 Latvia Phone 371 67426137 Fax 371 67428101 http www biosan lv Edition 4 6 02 December of 2018...

Отзывы: