11
6. Specifications
The unit is designed for operation in cold rooms and closed laboratory rooms at am-
bient temperature from +4°C to +40°C in a non-condensing atmosphere and maximum rel-
ative humidity 80% for temperatures up to 31°C decreasing linearly to 50% relative humidity
at 40°C.
Biosan is committed to a continuous programme of improvement and reserves the
right to alter design and specifications of the equipment without additional notice.
LMC-3000
LMC-4200R
Speed setting
range
In RPM
100
– 3000 RPM
100
– 4200 RPM
In RCF
1
10
– 1610 g
10
– 3160 g
Speed increment
100 RPM or 10 g
Timer setting
1
– 90 min
Timer increment
1 min
Acceleration,
RPM/s
Slow mode
50
50
Normal mode
75
75
Fast mode
2
135
150
Deceleration,
RPM/s
Brakes off
2
30
30
Slow mode
8
8
Normal mode
45
50
Fast mode
2
135
150
Temperature setting range
-
-
10 °C … +25 °C
Stable temperature
maintenance range
-
25 °C below ambient …
+25 °C
Temperature setting resolution
-
1 °C
Rotor imbalance automatic diag-
nostics
Emergency stop, dis
play indication “IMBALANCE”
Rotation direction
Counterclockwise
Display
2x16 characters, LCD
2x24 characters, LCD
Noise level
≤ 60 dBA
≤ 65 dBA
Chamber diameter
340 mm
360 mm
Dimensions
495x420x235 mm
635x580x335 mm
Power consumption
110 W (
V.6AD
);
120 W (
V.6AE
)
990 W
Operational voltage
230 V, 50/60 Hz (
V.6AD
);
120 V, 50/60 Hz (
V.6AE
)
230 V, 50/60 Hz
Weight
3
11.8 kg
56 kg
1
Relative Centrifugal Force
2
Depending on the load
3
Accurate within ±10%
Содержание LMC-3000
Страница 1: ......