4
DURING OPERATION
Do not centrifuge flammable or chemically active substances. If such liquids are
spilled on the rotor or rotor chamber, the centrifuge must be cleaned with a moist
cloth and a mild soap solution.
Do not use rotors with visible signs of corrosion, wear or mechanical damage.
Do not use the rotor without fixation nut (see figure below). Fixation procedure is
described in
4.4
.
Do not fill in the tubes after they have been inserted in the rotor.
Do not use glass tubes or tubes unsuitable for centrifugation.
Select the correct type of rotor. Some rotors have limited maximum speed (see figure
below). Selection procedure is described in
5.4.3
.
Do not leave the operating unit unattended.
Do not operate the unit in environments with aggressive or explosive chemical mix-
tures. Please contact manufacturer for possible operation of the unit in specific at-
mospheres.
Do not operate the unit if it is faulty or has been installed incorrectly.
Do not use outside laboratory rooms.
Do not place a load exceeding the maximum load value mentioned in the
Specifica-
tions
section of this manual.
BIOLOGICAL SAFETY
According to EN 61010-2-20, a centrifuge without a lid gasket is not considered a
biologically safe system and therefore cannot be used for centrifuging hazardous ma-
terials contaminated with toxic, radioactive or pathogenic microorganisms.
The user is responsible to carry out appropriate decontamination if hazardous mate-
rial spills on or penetrates into the equipment.
Rotor label