5.  Operation 

 

Recommendations during operation

 

 

Use  even  numbers  of  tubes  arranged  symmetrically  (one  opposite  another)  when 
loading to give the unit even balance during operation. The opposite tubes must be 
filled up equally. 

 

Centrifuge rotors must not be filled over the volume specified by the manufacturer. 

 

Do not fill in the tubes after they have been inserted in the rotor. 

 

Do not use glass tubes or tubes unsuitable for centrifugation. 

 

Rotor must always be fixed securely. Stop the operation immediately by pressing and 
holding the 

RUN/STOP

 

►/■ key for more than 2 seconds if any unusual noise occurs 

during acceleration, which can be due to improper rotor fixation. 

 

Since  some  plastic  tubes  and  microtest  plates  can  be  damaged  at  higher  speeds, 
maximum speed is limited for some rotors. Before centrifugation, select the type of 
installed rotor on the display, see 

5.4.3

 

Figure 4. LMC-3000 control panel 

 

Figure 5. LMC-4200R control panel 

5.1. 

Check  the  power  cable  for  any  signs  of  damage.  Connect  the  power  cable  to  a 
properly grounded mains socket. Set the power switch on the side to position 

I

 (on). 

 

Caution!

 

In model 

LMC-4200R

, cooling system requires some time for start-

up. After powering up, wait for about 4 minutes before centrifugation. 

5.2. 

The centrifuge turns on. The following readouts appear on the display: 

Acceleration and deceleration modes (fig. 4/2 or 5/2);  

Rotor selection (fig. 4/1 or 5/1); 

Set rotor speed in RPM or RCF

1

 (fig. 4/6 or 5/6); 

Set chamber temperature, in degrees Celsius (only 

LMC-4200R

, fig. 5/7); 

Status icons of rotor, ■ stopped or ► running (fig. 4/1 or 5/1, first symbol); 

Status icons of lid,   closed or 

 open (fig. 4/1 or 5/1, second symbol); 

Set time of centrifugation, in minutes and seconds (fig. 4/4 or 5/4); 

Actual rotor speed, in RPM or RCF (fig. 4/6 or 5/6); 

Actual chamber temperature, in degrees Celsius (only 

LMC-4200R

, fig. 5/8). 

 

1

 Revolutions per minute and Relative Centrifugal Force 

Содержание LMC-3000

Страница 1: ......

Страница 2: ...6 Specifications 11 7 Ordering information 12 8 Care and maintenance 13 9 Warranty 14 10 EU Declaration of conformity 15 1 About this edition of user instructions The current edition of the user inst...

Страница 3: ...hin a 300 mm area around the device during the centrifuge operation Use only original parts and accessories provided by manufacturer for this product Before using any cleaning or decontamination metho...

Страница 4: ...4 3 Do not leave the operating unit unattended Do not operate the unit in environments with aggressive or explosive chemical mix tures Please contact manufacturer for possible operation of the unit i...

Страница 5: ...pounds etc and material itself blood components CSF and other thermolabile materials Temperature control absence at this stage causes unpredictable results FEATURES User friendly centrifugation parame...

Страница 6: ...ench for rotor fixation 1 pcs Emergency opening tool for LMC 4200R 1 pce Operating manual declaration of conformity 1 copy 4 2 2 Optional accessories R 6 rotor on request R 6P rotor on request R 12 10...

Страница 7: ...wear or corrosion and replace if necessary Hold the rotor with one hand and using the supplied wrench for rotor replacement 13 mm turn the fixation nut fig 1 1 counter clockwise to release the rotor...

Страница 8: ...tor on the display see 5 4 3 Figure 4 LMC 3000 control panel Figure 5 LMC 4200R control panel 5 1 Check the power cable for any signs of damage Connect the power cable to a properly grounded mains soc...

Страница 9: ...eration speed values can be found in the Specifications section This information can also be found on the lid of the unit 5 4 3 Installed rotor fig 4 3 or 5 3 Select the rotor that is currently instal...

Страница 10: ...leration mode press and hold RUN STOP key for more than 2 seconds 5 9 Press the Open key and open the lid by lifting it upwards with your hand It is possible to unlock and open the lid only when the r...

Страница 11: ...Timer increment 1 min Acceleration RPM s Slow mode 50 50 Normal mode 75 75 Fast mode2 135 150 Deceleration RPM s Brakes off2 30 30 Slow mode 8 8 Normal mode 45 50 Fast mode2 135 150 Temperature settin...

Страница 12: ...0208 BK R 24 10 For 24 round bottom centrifuge tubes of 10 15 ml xH 16x90 mm N A 4000 RPM 2860g BS 010212 JK R 2 For 2 standard 96 well microplates LxWxHmax 128x85 6x45 mm 2000 RPM 560g 2000 RPM 560g...

Страница 13: ...n of the unit 8 3 2 Disinfect the unit after operation by cleaning the parts inside the centrifuge chamber The rotor and other accessories are autoclavable 120 20 min 8 3 3 For LMC 4200R Wipe away any...

Страница 14: ...200R the Premium class model one year of extended warranty is availa ble free of charge after registration during 6 months from the date of sale Online registration form can be found in section Warran...

Страница 15: ...EU Applicable Standards LVS EN 61326 1 2013 Electrical equipment for measurement control and laboratory use EMC requirements General requirements LVS EN 61010 1 2011 Safety requirements for electrical...

Страница 16: ...Biosan SIA Ratsupites 7 build 2 Riga LV 1067 Latvia Phone 371 67426137 Fax 371 67428101 http www biosan lv Edition 4 6 02 December of 2018...

Отзывы: