6
3. General information
Biosan Labaqua is Ultrapure water purification system which requires external water
tank that use tap water as feed water (tap water system). There are three models available:
Labaqua Trace, Labaqua HPLC and Labaqua Bio.
Water purification system Biosan Labaqua produces pure water that complies with
ISO 3696 Grade II and Grade I water requirements.
Pure (ISO 3696 Grade II) water applications include, but are not limited to:
Feed for laboratory equipment (washing machines, clinical analyzers, humidifiers,
autoclaves, hydrogen gas generators);
Manufacturing of chemical and biochemical reagents;
Buffer preparation;
Microbiological media preparation;
In some cases - sensitive analytical techniques (e. g. atomic absorption, ICP-OES);
Wet chemistry;
Spectrophotometry.
Ultrapure (ISO 3696 Grade I) water applications include but are not limited to:
High sensitivity analytical techniques (ICP-MS);
High performance liquid chromatography;
TOC analysis
Molecular biology;
Cell culture.
Table 1. Labaqua model comparison
Model
Labaqua Trace Labaqua HPLC
Labaqua Bio
Grade I water resistivity
18.2 MΩ x cm
Grade I water conductivity
0.055 μS/cm
Grade II water conductivity
0.1 μS/cm
TOC
< 10 ppb
< 2 ppb
RNase
-
-
< 0.01 ng/mL
DNase
-
-
< 4 pg/μL
Bacteria
< 1 CFU/mL
< 0.01 CFU/mL
Endotoxins
0.15 EU/mL
0.001 EU/mL