4.5.
Feedwater supply connector and drain should be within 3 meters from the unit loca-
tion.
Attention
Please note that if the pre-filter clogs prematurely, such malfunction
will be covered by warranty only of the service department receives
photographic proof of the particle filter installed in the feedwater ac-
cess pipe.
4.6.
Requirements for the installation site. Ensure that the unit is placed on a solid, level
surface not less than 320x560 mm, which is able to support its weight and the weight
of water inside the unit. The storage tank Pro requires at least 300x300 mm free
space. The tank during operation weighs up to 40 kg. The tank can be placed below
the unit. Ensure that all cables, tubes abd the power switch on the rear panel of the
unit are easily reachable.
4.7.
Requirements for the environment. The water purification unit is intended for indoor
use only. Ensure that the site is maintained under the following conditions:
-
Temperature ........................................................................
15 to 30 °C (59 to 86 °F)
-
Humidity ...................... 20% to 80% relative humidity, non-condensing atmosphere.
4.8.
Water leakage safety:
-
Ensure that no water tubes are bent.
-
Ensure that all connections are watertight.
-
Feedwater access and drain connections must be no farther than 3 m from the unit.
-
Water access connector is ½" male NPTF.
Attention!
Ensure that the storage tank is connected to the drain with a tube
from the OVERFLOW fitting. Ensure that the drain is lower than the
OVERFLOW fitting. This prevents water leakage in case of water
level sensor malfunction.
Attention!
Ensure that the 1 μm particle filter is installed on the feedwater
access pipe.
Failure to do so will lead to pre-filter clogging in the unit
and water flow restriction.