1. Safety precautions

The following symbols mean

:

Caution

!

 

Make sure you have fully read and understood the present Manual 

before using the equipment. Please pay special attention to 
sections marked by this symbol.

GENERAL SAFETY

·

Use only as specified in  the operating instructions provided.

·

The unit should be saved from shocks or 

falling

.

·

The unit must be stored and transported in a horizontal position (see package 

label).

·

After transportation or storage keep the unit under room temperature for 2-3 hrs 

before connecting it to the electric circuit.

·

Use only cleaning and decontamination methods recommended by the 

manufacturer.

·

Do not make modifications to the design of the unit.

ELECTRICAL SAFETY

·

Connect only to a power supply with voltage corresponding to that on the serial 

number label.

 

·

Use only the external power supply unit provided with this product.

·

Ensure that the external power supply connector 

is

 easily accessible during use.

·

Disconnect the unit from the electric circuit before moving.

·

Turn off the unit by disconnecting the external power supply from the power 

socket.

·

If liquid penetrates into the unit, disconnect it from the external power supply unit 

and have it checked by a repair and maintenance technician.

DURING OPERATION 

·

Do not operate the unit in environments with aggressive or explosive chemical 

mixtures.

Содержание Bio RS-24

Страница 1: ...Multi Bio RS 24 Programmable rotator for version V 3AW Operating Manual Certificate Medical Biological Research Technologies...

Страница 2: ...Contents 1 Safety precautions 2 General information 3 Getting started 4 Operation 5 Program setting 6 Specifications 7 Maintenance 8 Warranty and Claims 9 Declaration of Conformity Page 2...

Страница 3: ...ing and decontamination methods recommended by the manufacturer Do not make modifications to the design of the unit ELECTRICAL SAFETY Connect only to a power supply with voltage corresponding to that...

Страница 4: ...o run on the borders of reciprocal motion segment It is available only when the reciprocal motion is ON Do not operate the unit if it is faulty or has been installed incorrectly Do not use outside lab...

Страница 5: ...vide increased methodical means to researchers working in the field of modern molecular and cell biology and the developing biodiagnostics technology based on the use of magnetic particles for which u...

Страница 6: ...1 copy Optional accessories PRS 5 12 platform v on request PRS 10 platform w on request PRSC 22 platform x on request PRSC 10 platform y on request PRS 1DP platform z on request 3 3 Set up place the...

Страница 7: ...set and the timer indicator Fig 1 2 shows 0 00 pressing the Run Stop key will cause continuous operation of the rotator until the Run Stop key is pressed again 4 7 If the operation time is set the pla...

Страница 8: ...perations the microprocessor saves the last parameter changes as the working program automatically 5 4 The countdown timer is used to control the operation time The timer can be set for the period fro...

Страница 9: ...time of Reciprocal motion is shorter or equal to the set time of Vibro pause mode the Reciprocal motion mode will be omitted Orbital Pause 5 5 5 Reciprocal Rotation Set the speed 1 100 rpm for Orbital...

Страница 10: ...r performs soft vibration Vibro motion However there is a mode for hard vibration To perform hard vibration set the turning angle of Reciprocal motion to 90 and the angle of Vibro type motion to 1 Har...

Страница 11: ...ation mode Tilt angle range 1 90 increment 1 Time setting range 0 250 sec 6 3 Vibro pause mode Tilt angle range 0 5 increment 1 Time setting range 0 5 sec 6 4 Digital time setting 1 min 24 hours or no...

Страница 12: ...continuous programme of improvement and reserves the right to alter design and specifications of the equipment without additional notice Up to 100 g for platform with rubber clamps or 200 g for platf...

Страница 13: ...your local Biosan representative 7 2 All maintenance and repair operations must be performed only by qualified and specially trained personnel 7 3 Standard ethanol 75 or other cleaning agents recomme...

Страница 14: ...istributor to check availability of extended warranty 8 3 If any manufacturing defects are discovered by the Customer an unsatisfactory equipment claim shall be compiled certified and sent to the loca...

Страница 15: ...Version 3 03 October 2012 9 Declaration of Conformity...

Страница 16: ......

Отзывы: