Step 8. Hold the plate by your hand and press down the tube of Alignment Marker
tightly into the well by thumb
Step 9.
Allocate your samples (sample volume ≥ 20 μ
l) Spin down and make sure the
sample is in the bottom of tube and no air bubble appears
※
The lid of 0.2 ml tube that placed at A1- A12 position must be removed/cutoff
Step 10.
Click “Park”, to move the holder back to the park position and click “Close
Sample Door”
Step 11. After unpacking cartridge, open the cartridge door and insert multi-channel
cartridge
※
L-shape connector of cartridge should follow the L-shape guiding groove inside
the instrument
Step 12. Press the cartridge door to close it
Step 13.
Click “Latch” The Cartridge Information will be displayed on the screen