Oxy9Wave
Vet
Operation manual
Ver. 1.52 5.Alarms/Messages 38
5-4. System messages
The chart below alphabetically lists all system messages displayed on the LCD screen
(the cause of the message and action(s) to be taken are also shown)
The operator should thoroughly familiarize himself/herself with this information before using the
oximeter for patient monitoring.
ARTIFACT -
The SPO
₂
signal is the patient
’
s motion artifact and noise.
Check Probe -
SPO
₂
data continues to be displayed, but the quality of the signal is questionable.
Check the patient and the probe.
Lead Fault
–
The multisite probe was detached from the patient. Check the probe. The factory default for this
alarm is MESSAGE ALARM.
Poor Signal
–
The SPO
₂
signal is too low. No SPO
₂
data is displayed. This can be due to low patient pulse,
patient motion, or some other interference. Check the patient and the probe.
Pulse Search
–
Detection by the monitor of a repeatable pulse has ceased. Check the patient and the probe site.
NOTE
: No SPO
₂
data is displayed. One of the following conditions is indicated:
◼
Defective or damaged probe
◼
Defective or damaged cable
◼
Probe is detached from the patient.
◼
Detection of repeatable pulse has ceased.
◼
Check the probe and cable: reposition or replace as needed.
NOTE :
When
‘
Artifact Signal
’
is displayed, measured values are displayed in gray.
◼
These indicate the most recently measured value among those measured under normal
conditions
Содержание Oxy9Wave Vet
Страница 1: ...Oxy9Wave Vet Operation manual 1 Oxy9Wave Vet Operation Manual Ver 1 52 www ebionet com...
Страница 15: ...Oxy9Wave Vet Operation manual Ver 1 52 2 Product description 15 Slide View Back View...
Страница 20: ...Oxy9Wave Vet Operation manual Ver 1 52 3 Setup 20 3 4 CONNECTING a Clamp...
Страница 47: ...Oxy9Wave Vet Operation manual Ver 1 52 47...