19
www.bioneer.com
BQ-042-101-13
Revision :0 (2018-07-05)
AccuLoader
™
9)
Scan the first sample tube with the barcode scanner and dispense the assigned volume of samples
into the designated wells using a pipette.
Repeat the barcode scanning and sample dispensation steps as many times as the assigned pool
size for the sample tubes.
<Figure 12. Sample dispensation>
10)
All completed status will be displayed.
When all dispensation has been completed,
“
Remain Tube: 0
”
and
“
Load Completed!
”
will be
displayed at the bottom and the upper right side, respectively.
Tap the END button for initialization of the instrument to open the contamination shield cover.